Performance Test Record - AC6803A
Test Description
Minimum Spec.
Results
Maximum Spec.
AC Voltage Programming and Readback Accuracy
135 Vrms @ 60Hz
134.595
______ V
135.405
Front Panel Measurement
Vrms - 0.975 V
______ V
Vrms + 0.975 V
135 Vrms @ 500Hz
134.595
______ V
135.405
Front Panel Measurement
Vrms - 1.845 V
______ V
Vrms + 1.845 V
270 Vrms @ 60Hz
269.325
______ V
270.675
Front Panel Measurement
Vrms - 1.95 V
______ V
Vrms + 1.95V
270 Vrms @ 500Hz
269.325
______ V
270.675
Front Panel Measurement
Vrms - 3.69 V
______ V
Vrms + 3.69V
Voltage Load Regulation
CV Load Effect Low Range @ 60 Hz
Vout - 0.15 V
______ V
Vout + 0.15 V
CV Load Effect Low Range @ 500 Hz
Vout - 0.50 V
______ V
Vout + 0.50 V
CV Load Effect High Range @ 60 Hz
Vout - 0.30 V
______ V
Vout + 0.30 V
CV Load Effect High Range @ 500 Hz
Vout - 1.00 V
______ V
Vout + 1.00 V
RMS Current Measurement Accuracy
Low Range Measurement @ 60 Hz
Iout – 180 mA
______ A
Iout + 180 mA
Low Range Measurement @ 500 Hz
Iout – 300 mA
______ A
Iout + 300 mA
High Range Measurement @ 60 Hz
Iout – 90 mA
______ A
Iout + 90 mA
High Range Measurement @ 500 Hz
Iout – 150 mA
______ A
Iout + 150 mA
Frequency Accuracy
Program 500 Hz
499.
9
Hz
______ Hz
500.
1
Hz
DC Voltage Programming and Readback Accuracy
+190 VDC Output
189.430 VDC
______ VDC
190.570
Front Panel Measurement
Vout - 1.25 VDC
______ VDC
Vout + 1.25 VDC
+380 VDC Output
379.050 VDC
______ VDC
380.950
Front Panel Measurement
Vout - 2.50 VDC
______ VDC
Vout + 2.50 VDC
DC Current Readback Accuracy
Low Range Measurement
Iout – 160 mA
______ A
Iout + 160 mA
High Range Measurement
Iout – 80 mA
______ A
Iout + 80 mA
Keysight AC6800 Series Operating and Service Guide
183