OUTPut:PON:STATe RST|RCL0|AUTO
OUTPut:PON:STATe?
Sets the power-on output state.
Parameter
Typical Return
RST|RCL0|AUTO
RST, RCL0, or AUTO
Set the power-on state to the *RST state: OUTP:PON:STAT RST
–
RST
specifies that the instrument powers on in a state equivalent to *RST.
–
RCL0
loads the state stored in state storage memory location 0.
–
AUTO
specifies that the instrument powers on in the power-down state. The output state will always be off.
– This parameter is not affected by
or
– This parameter is saved in nonvolatile memory.
OUTPut:PROTection:CLEar
Resets the latched protection. This clears the latched protection status that disables the output when a protection
condition occurs.
Parameter
Typical Return
(none)
(none)
Clears the latched protection status: OUTP:PROT:CLE
– All conditions that generate the fault must be removed before the latched status can be cleared. The output
remains in the OFF state after the fault condition is cleared.
OUTPut:PROTection:WDOG[:STATe] ON|1|OFF|0
OUTPut:PROTection:WDOG[:STATe]?
Enables or disables the I/O watchdog timer.
Parameter
Typical Return
0|OFF|1|ON
0 (OFF) or 1 (ON)
Enables the watchdog timer protection: OUTP:PROT:WDOG ON
– When the watchdog timer is enabled, the output will be disabled if there is no I/O activity on any remote interface
within the time period specified by OUTput:PROTection:WDOG:DELay. The output is latched off but the
programmed output state is not changed.
– The watchdog timer function is NOT reset by front-panel activity; the output will still shut down after the time
period has elapsed.
– The
command sets this parameter to 0.
136
Keysight AC6800 Series Operating and Service Guide