Current High Range
Use the same setup as used for Current Low Range calibration.
Step
Description
Front Panel
SCPI
1
Select
High
range.
[Menu] > System > Admin > Cal > Current > High_
Range
CAL:CURR:OFFS 2
2
Disconnect the load and shunt.
3
Go to next step.
Select
Next
.
CAL:LEV P1
*OPC?
4
Wait five minutes.
5
Go to next step.
Select
Next
.
6
Reconnect the load and shunt.
7
Select
Next
.
8
Enter a 0 (zero).
(no action required)
CAL:DATA 0
9
Select
Next
.
10
Set the Keysight 6060B to Constant Current Mode and set the load current to 90% of the value from the table
below for the model being calibrated. For example, use 1.8 A for the AC6801A.
11
Go to next step.
Select
Next
.
CAL:LEV P2
*OPC?
12
Wait five minutes.
13
Select
Next
.
14
Calculate the shunt
current (I=V/R) and
enter the data.
Enter the computed current from the external DMM
measurement and Load resistance and press
[Select]
or
[Enter]
. Select
Next
or press
Back
to finish the calibration
step.
CAL:DATA <
data
>
Current Table, High Range
AC6801A AC6802A AC6803A AC6804A
High Range
2.0 A
4.0 A
8.0 A
16.0 A