FIGURE 3-7. WEB INTERFACE OPERATE INSTRUMENT PAGE
3.3.6.5.1 CHANGING THE OUTPUT [-1.2K MODELS ONLY]
The Output dialog box opens when the OPERATE INSTRUMENT page is first opened, or when
OUTPUT
is clicked (see Figure 3-7). To set output voltage or current, enter the values and click
SET. Voltage and current settings are not allowed to exceed either a) the virtual model setting or
b) 80% of the protection value. Attempts to set voltage or current beyond these limits will not be
accepted. To turn the output on or off, click the ON/OFF button within the simulated front panel.
3.3.6.5.2 CHANGING PROTECTION [-1.2K MODELS ONLY]
Click
PROTECTION
to open the Protection dialog box (Figure 3-8). To set the overvoltage and
overcurrent protection values, enter the values and click
Set
. See PAR. 3.2.8 for details about
how overvoltage and overcurrent protection function in the KLP power supply.
FIGURE 3-8. PROTECTION DIALOG BOX
3.3.6.5.3 USING SAVE AND RECALL [-1.2K MODELS ONLY]
The KLP Power Supply contains 40 memory locations that can be used to store either active
settings or a set of user-determined settings. Values are stored in the nonvolatile memory and
are retained when the unit is turned off. The stored settings can then be recalled using this fea-