How do I view/change the virtual model settings?

 See PAR. 3.2.6 for description of virtual model. 

If status shows SET, exit set-

point mode (tap either knob once).

 Use thin tool to press FUNCTION once; status reads VIRT. VOLTS and AMPS displays show the 

maximum voltage and current of the present virtual model. 

If no change

, exit function mode (see above). To change, press DC OUT-

PUT once. Status shows PASS (password). Enter password 

(see above)

 and tap DC OUTPUT to accept. When password is 

accepted, status reads VIRT, VOLTS and AMPS displays blink. To change setting rotate Voltage and Current knobs until VOLTS and 
AMPS displays show desired virtual model settings. (Press and hold knob in while turning for fine adjustment.) Press DC OUTPUT to 
accept virtual model settings. Unit displays DONE, then resets and repeats power on sequence.

How do I view/change overvoltage and overcurrent protection limits?

 See PAR.3.2.6 for description of protection limits. 

If status 

reads SET, tap either knob to exit setpoint mode

. To view protection limits, use a thin tool to press and hold PROTECT until status 

reads PROT, VOLTS and AMPS displays show the protection limits; to change limits, while continuing to hold PROTECT in, adjust 
OVP and OCP limits using the Voltage and Current knobs. When PROTECT released, values showing in VOLTS and AMPS displays 
are the new protection limits.

REMOTE ANALOG OPERATING MODE INSTRUCTIONS

How do I enable analog operating mode?

 Connect ANALOG_CTRL (pin 12) to GROUND (pin 9, 11, 13, or 15) of analog port prior 

to power-up to enable remote analog programming for that session. Remote analog control is indicated by "aREM" in status display. 

The status of ANALOG_CTRL is sensed at power-up only

. Enabling analog operating mode disables the local operating controls on 

the front panel. (see PAR. 3.4).

How do I enable or disable the output?

 Connect REM_INH (pin 8) to GROUND (pin 9, 11, 13, or 15) of analog port to disable 

power supply output, open the connection to enable the output (see PAR. 3.4.1). The DC OUTPUT control on the front panel remains 
active in remote analog programming mode. 

How do I adjust voltage and current setpoints using an external voltage source?

 Set analog I/O DIP switch positions 1 and 2 up 

(factory default position) to use external voltage control of voltage and current, respectively. Full-scale programming voltage is factory 
set at 10V; to alter full-scale value see PAR. 3.2.13.6. Apply a 0-10V signal from Vref (pin 1) and Cref (pin 3) to GROUND (pin 9, 11, 
13, or 15) of analog port to adjust the setpoint values for output voltage and current proportionally between minimum and maximum 
virtual model limits. 

Programming input pins are only active when analog port is enabled.

 (See PAR. 3.4.3 for details.)

How do I adjust voltage and current setpoints using an external resistance?

 Set analog I/O DIP switch positions 1 and 2 down 

to use external resistance for remote analog programming of voltage and current, respectively. Full-scale programming resistance is 
factory set at 10KOhms; to alter full-scale value see PAR. 3.2.13.6 and use 1000 ohms per volt scale factor. Apply a 0-10KOhm resis-
tance from Vref (pin 1) and Cref (pin 3) to GROUND (pin 9, 11, 13, or 15) of analog port to adjust the setpoint values for output volt-
age and current proportionally between minimum and maximum virtual model limits. 

Programming inputs are only active when analog 

port is enabled. 

(See PAR. 3.4.2 for details.)

How do I Read back operating voltage and current?

 Analog signals proportional to actual output voltage and current levels are 

available at VOLT_RBACK (pin 6) and CURR_RBACK (pin 7) of analog I/O port. Full-scale signal level is the same as that for full 
scale programming. Readback signals are always available at analog port, even if analog programming is not enabled. 

How do I exit remote analog programming mode?

 Simultaneously depressing and releasing both Voltage and Current knobs 

forces the unit to exit remote analog programming mode and return to local operating mode; this can also be done by sending any 
command or query via the remote digital inputs. To restore remote analog control cycle source power OFF for five seconds then 
restart.

REMOTE DIGITAL OPERATING MODE INSTRUCTIONS

How do I enable digital operating mode?

 Sending any digital command or query via either the GPIB, LAN or RS 232 port immedi-

ately sets the unit in remote digital operating mode and disables the local operating controls on the front panel. For details on com-
mand formatting and communication protocols see PAR. 3.3.

How do I exit remote digital programming mode? 

Simultaneously depressing and releasing both Voltage and Current knobs will 

force the unit to exit remote digital programming mode and return to local operating mode. Remote digital control is reactivated upon 
transmission of any digital command or query by the control (host) computer.

How do I view/change the GPIB address?

 The factory default GPIB address is 6. Using a thin tool, press FUNCTION once; status 

reads VIRT. Rotate Voltage or Current knob until status reads ADDR; the active GPIB address shows in AMPS display. To exit, 
rotate Voltage or Current knob till status reads EXIT then tap DC OUTPUT. To change, tap DC OUTPUT: AMPS display blinks. 
Rotate Current knob to change address and tap DC OUTPUT to accept; display flashes DONE then ADDR. See above to exit.

How do I view/change RS 232 baud rate? [-1200 Models Only]

 The factory default RS 232 baud rate is 38400. Using a thin tool, 

press FUNCTION once; status reads VIRT. Rotate Voltage or Current knob until status reads BAUD; AMPS display shows active 
baud rate x 1000 (e.g., 38400 is displayed as 38.4). To exit, rotate Voltage or Current knob till status reads EXIT then tap DC OUT-
PUT. To change, tap DC OUTPUT: AMPS display blinks. Rotate Current knob to change baud rate and tap DC OUTPUT to accept; 
display flashes DONE then BAUD. See above to exit. See PAR. 3.3.1.2 for additional RS 232 parameters. 

How do I set up the LAN interface?

 

[-1.2K Models Only]

 See PAR. 3.2.12 and refer to KLP Developer’s Guide.

How do I operate the KLP using web pages via the LAN interface?

 

[-1.2K Models Only]

 See PAR. 3.3.6 and refer to KLP Devel-

oper’s Guide. Use the PSfind program (on CD supplied) to locate KLP(s) functioning on the LAN. Highlight the unit and click Launch 
Web Page. Click OPERATE INSTRUMENT. Simulated front panel shows status of voltage, current, output on/off, and operating 
mode (CV or CC). Operating functions supported from Operate Instrument page: Set output on/off and output voltage/current (PAR. 
3.3.6.5.1), overvoltage/overcurrent protection (PAR. 3.3.6.5.2), save/recall (PAR. 3.3.6.5.3), reset (PAR. 3.3.6.5.4), list (PAR. 
3.3.6.5.5) and virtual model (PAR. 3.3.6.5.6). Click CONFIGURE LAN to change LAN addresses or set up LAN passwords (PAR. 
3.3.6.3). Click ID INSTRUMENT to flash physical LAN indicator on KLP.

Summary of Contents for KLP 10-150-1.2K

Page 1: ...REV 1 KLP 75 33 1200 REV 25 KLP 75 33 1 2K REV 1 KLP 150 16 1200 REV 22 KLP 150 16 1 2K REV 1 KLP 300 8 1200 REV 14 KLP 300 8 1 2K REV 1 KLP 600 4 1200 REV 3 KLP 600 4 1 2K REV 1 2 A Change Page may b...

Page 2: ......

Page 3: ...USA Importer s Name and Address Type of Equipment Component Power Supply Model No PRODUCT MODEL NUMBER Year of Manufacture I the undersigned declare that the product specified above when used in conj...

Page 4: ...wire grounded mains connection See wiring section of this manual for complete electrical wiring instructions EN61010 1 2001 Cl 6 10 1 4 This power supply has secondary output circuits that are conside...

Page 5: ...orming any servicing The PROTECTIVE EARTHING CONDUCTOR TERMINAL must be properly con nected prior to application of source power to the product see instructions on instal lation herein in order to ens...

Page 6: ......

Page 7: ...Controls and Indicators 2 1 2 2 2 Rear Panel Connectors and Switches 2 1 2 3 Source Power Requirements 2 6 2 4 Cooling 2 6 2 5 Preliminary Operational Check 2 6 2 6 Installation 2 7 2 6 1 Rack Mountin...

Page 8: ...actory Default Settings 3 18 3 3 1 1 GPIB Factory Defaults 3 18 3 3 1 2 RS 232 Factory Defaults 1200 Models Only 3 18 3 3 1 3 LAN Factory Defaults 1 2K Models Only 3 18 3 3 2 Additional Functions Not...

Page 9: ...4 3 3 External Calibration Local 4 3 4 3 4 Calibration Exit Local 4 4 4 4 Calibration using VXI plug play Driver 4 4 4 4 1 Voltage Calibration VXI plug play Driver Demo 4 6 4 4 2 Current Calibration...

Page 10: ...c IP Address 3 21 3 4 Web Interface Home Page Unit Description 3 21 3 5 Web Interface Configure Instrument Page 3 22 3 6 Web Interface Configure LAN Page 3 24 3 7 Web Interface Operate Instrument Page...

Page 11: ...unctions 2 2 2 3 IEEE 488 Port Connector J4 Pin Assignments 2 4 2 4 RS232 Port Connector J3 Pin Assignments 1200 Models Only 2 4 2 5 LAN Port Connector J7 Pin Assignments 1 2K Models Only 2 5 2 6 Cont...

Page 12: ...vi KLP031308 FIGURE 1 1 KLP SERIES POWER SUPPLY...

Page 13: ...um current at rated voltage and rated current and maximum voltage at rated current parameters define the virtual models available within the limit of 1200 Watts of output power see Figure 1 2 KLP powe...

Page 14: ...V 1 6A 0 120A 10V 120A 75 mV 82 KLP 36 60 3 0 36V 33 3A 36V 0 8A 0 60A 20V 60A 125 mV 83 KLP 75 33 3 0 75V 16A 75V 0 4A 0 33 3A 36V 33 3A 125 mV 84 KLP 150 16 3 0 150V 8A 150V 0 2A 0 16A 75V 16A 125 m...

Page 15: ...current mininimum to 100 of rated current No minimum load required See Table 1 1 for minimum programmable current Source effect voltage 0 01 Emax Over full source range current 0 01 Imax Load effect v...

Page 16: ...36V models Nickel plated copper busbars Provision for safety covers 75 600V models Shocksafe Euroblock Analog programming port 15 pin D sub Digital programming ports primary Standard GPIB connector I...

Page 17: ...3 2 13 5 1 Composite status flag Isolated form C contacts Programmable functions PAR 3 2 13 3 PROGRAMMING CHARACTERISTICS DIGITAL Supported Interfaces 1200 Models GPIB and RS 232 SCPI command set for...

Page 18: ...1 6 KLP 031308 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2 R DWG MECH 3010266...

Page 19: ...KLP 031308 1 7 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...

Page 20: ...ds or via the LAN port using SCPI commands or web pages Most features available in local mode can be accessed in remote mode through dig ital programming as well as some that are only available via di...

Page 21: ...00 memory locations for programming sequences that can program the KLP to produce a variety of sequential output operations Each memory location accommodates a value for voltage and current and a dwel...

Page 22: ...require the use of multiple KLP 1200 Model power supplies in either series for higher voltage or parallel for redundancy or higher current the master slave feature allows a single power supply the ma...

Page 23: ...SNC 488 1 IEEE 488 GPIB Cable 2m long Connect KLP Power Supply to GPIB bus SNC 488 2 IEEE 488 GPIB Cable 4m long Connect KLP Power Supply to GPIB bus SNC 488 4 Chassis Slide 2 required per power supp...

Page 24: ......

Page 25: ...utton switch Rotate to set output voltage PAR 3 2 7 and overvoltage limit PAR 3 2 8 Also used to enter SET mode of voltage programming PAR 3 2 7 2 change GPIB address PAR 3 2 9 change RS232 baud rate...

Page 26: ...n without blinking LED not lit indicates failure FUNCTION momentary switch Used to set user determined limits on voltage and current virtual model PAR 3 2 6 change GPIB address PAR 3 2 9 change RS232...

Page 27: ...KLP HV 031308 2 3 FIGURE 2 2 KLP SERIES REAR PANEL SWITCH AND CONNECTORS...

Page 28: ...common 22 GND Ground signal common 23 GND Ground signal common 24 LOGIC GND Logic Ground TABLE 2 4 RS232 PORT CONNECTOR J3 PIN ASSIGNMENTS 1200 MODELS ONLY PIN SIGNAL NAME FUNCTION NOTE A null modem c...

Page 29: ...position 1 See Table 2 2 Refer to PAR 3 4 3 for voltage programming and 3 4 2 for resistance programming 4 RELAY_COM Relay common 1 5 Reserved 6 VOLT_RBACK Analog signal which represents output volta...

Page 30: ...llation is advisable to ascertain whether the power supply has suffered damage resulting from shipping Refer to Figures 2 1 and 2 2 for location of operating controls and electrical connections Table...

Page 31: ...to enable the output Verify DC OUTPUT indicator lights 8 Compare the programmed output voltage value step 5 with the voltage reading of the DVM the difference between the two should not exceed 0 05 o...

Page 32: ...under worst case efficiency conditions at the lowest rated source voltage 100V a c and provides the necessary overload protection to the power supply s internal circuitry The user must provide a prope...

Page 33: ...ound point is entirely dependent upon the specific applica tion and its selection requires a combination of analysis good judgement and some amount of empirical testing If there is a choice in selecti...

Page 34: ...e power leads it cannot completely compensate for the undesirable output effects of the power lead inductance These lead impedances both power and sensing leads are especially important if the load is...

Page 35: ...d 3 2 11 2 7 7 PARALLEL REDUNDANT OPERATION Identical KLP power supply models may be connected in parallel in order to provide increased output current to a common load see Figure 2 5 This permits the...

Page 36: ...2 12 KLP HV 031308 FIGURE 2 4 SERIES CONNECTION USING REMOTE SENSING FIGURE 2 5 PARALLEL CONNECTIONS USING REMOTE SENSING 2 7 6...

Page 37: ...ired power lead voltage drops must be minimized Provide local noise decoupling capacitors across all sense wire termination points b The power supplies should be located as near to each other as possi...

Page 38: ...dback VOLT_RBACK and CURR_RBACK Twist the voltage readback signal VOLT_RBACK pin 6 Current readback signal CURR_RBACK pin 7 together with ground pin 9 11 13 or 15 Full scale readback volt ages are ide...

Page 39: ...lder Apple MAC computers with D sub serial port See Table 2 4 for RS 232 port pin assignments Connect the cable to the KLP at the RS 232 port located on the rear panel See Figure 2 2 Refer to PAR 3 2...

Page 40: ......

Page 41: ...ions are provided by a 4 character alpha numeric display 3 2 1 TURNING THE POWER SUPPLY ON 1 To turn the power supply on set POWER ON OFF circuit breaker 1 Figure 2 1 to ON Apply firm continuous press...

Page 42: ...med current is automatically set to be at least the minimum current actual value depends on model see Table 1 1 Since the output is off Constant Voltage CV mode indicator green LED and Con stant Curre...

Page 43: ...unit will accept Analog Remote commands see PAR 3 4 and front panel controls are disabled except that the DC OUTPUT ON OFF switch is operational Local mode can be restored by pressing the VOLTAGE and...

Page 44: ...LTAGE or CURRENT control The status display reads SET and the DC VOLTS and DC AMPERES displays show the stored setpoints To accept the displayed value tap the associated adjustment control again To ch...

Page 45: ...cessible from the front panel see PAR 3 3 2 should be checked to ensure that there are no settings outside the range permitted by the new maximum programmable voltage and current Storage of User progr...

Page 46: ...T Rotating the associated control will change the output voltage or current in real time only if the output is enabled If the unit is in constant voltage mode CV indicator lit the DC VOLTS display sho...

Page 47: ...ed value buffer is cleared and setpoints will default to zero volts and minimum current NOTE If the virtual model for the power supply is altered the last settings are reset to zero volts and minimum...

Page 48: ...ayed GPIB address The display briefly reads DONE then returns to ADDR 5 Rotate the VOLTAGE control until display reads EXIT then tap the DC OUTPUT switch to exit 3 2 10 CHANGING RS232 BAUD RATE 1200 M...

Page 49: ...ting the control Both connected units must be set the same either SERI or PARA Specify the total number of units in the configuration to 2 by rotating the CURRENT con trol Specify the unit number to b...

Page 50: ...2 11 3 OPERATING THE MASTER SLAVE CONFIGURATION 1200 MODELS ONLY Once the master is configured it searches for the slave specified Until the slave is set up prop erly the master status display will c...

Page 51: ...ther VOLTAGE or CURRENT control to see desired setting and press DC OUTPUT switch to save ON The KLP requests an IP address from the DHCP server AUTOIP set to OFF OFF The KLP uses the IP MASK GATE and...

Page 52: ...escribed above RTN Return to LAN menu 3 2 13 UTILITY FUNCTION The Utility Function provides additional system information or access to special functions These are described below 1 If the status displ...

Page 53: ...m the front panel to operate in either Fault Manual or List mode If FAULT mode the factory default is selected the relay energizes automatically upon detection of a user selected condition The selecti...

Page 54: ...ay determines the length of time that the unit must be operat ing in the new mode before energizing the relay The factory default delay time is set to 1 0 seconds to prevent false tripping of the rela...

Page 55: ...n active TTL high or low and 2 how the unit recovers from a fault The unit password is required to change these functions 1 Enter UTIL menu see PAR 3 2 13 steps 1 through 3 and rotate either VOLTAGE o...

Page 56: ...g a safe response output off but it is the user s responsibility to NOT attempt multiple consecutive restarts without eliminating the fault condition AUTOmatic When a fault configured as AUTO is detec...

Page 57: ...lay shows the fault The setting in the DC VOLTS display begins to blink Rotate either VOLTAGE or CURRENT control to cycle through the allowable settings Tap the DC OUTPUT switch to accept the setting...

Page 58: ...EFAULTS 1200 MODELS ONLY The factory default RS 232 settings are as follows refer to PAR 3 2 10 to change the RS 232 baud rate 1200 Models only Baud rate 38400 Parity None Data Bits 8 Stop Bits 1 Echo...

Page 59: ...e www kepco power com drivers Examples are provided in the KLP Developer s Guide 3 3 4 REMOTE PROGRAMMING USING THE LABVIEW G INSTRUMENT DRIVER The LabView G driver for KLP available for download at t...

Page 60: ...he left side of the window If you do not see your device in the Select a Unit window make sure it is turned on and con nected to the network then click the Search Again button To set a static IP addre...

Page 61: ...KLP is removed or lost while a browser window is open commu nication with the KLP can not be restored until the present browser window is closed and a new window is opened Although most current brows...

Page 62: ...elected This is especially important if more than one unit is found on the LAN 3 3 6 3 INSTRUMENT CONFIGURATION USING WEB INTERFACE 1 2K MODELS ONLY Click on CONFIGURE INSTRUMENT at the left to view t...

Page 63: ...shown on the CONFIGURE INSTRUMENT page The CONFIGURE LAN password restricts the ability to change the LAN configuration parameters shown on the CONFIGURE LAN page Figure 3 6 The factory default is th...

Page 64: ...lt and operation via the web is disabled When the fault is corrected click OK to close the alert window and resume operation via the web NOTE If the OPERATE INSTRUMENT password has been set see PAR 3...

Page 65: ...ick the ON OFF button within the simulated front panel 3 3 6 5 2 CHANGING PROTECTION 1 2K MODELS ONLY Click PROTECTION to open the Protection dialog box Figure 3 8 To set the overvoltage and overcurre...

Page 66: ...the modified settings To save the active settings Enter the desired location 1 to 40 in the LOCATION box and click Save State button to save the active settings Only the parameters shown in Figure 3...

Page 67: ...and are retained until reprogrammed the power supply is turned off a calibration is performed or a LIST CLEar command is received If a virtual model setting is changed and a programmed value is outsi...

Page 68: ...amp buttons however the total number of steps can not exceed 100 To execute the list enter the number of times the list is to be executed in the Loops box 0 is continuous set the output on click the O...

Page 69: ...to full scale see PAR 4 3 3 will cause a corresponding decrease in programming resolution With the exception of Vref and Cref all the other Analog I O Port functions are active regardless of the statu...

Page 70: ...or ON Local controls locked 3 Configure Analog I O Port J2 Figure 2 2 by referring to Table 2 7 and Figure 3 14 ground ing J1 pin 12 enables analog programming FIGURE 3 14 ANALOG PROGRAMMING OF OUTPU...

Page 71: ...ither OFF Local controls enabled or ON Local controls locked 3 Configure Analog I O Port J2 Figure 2 2 by referring to Table 2 7 and Figure 3 15 ground ing J1 pin 12 enables analog programming 4 Refer...

Page 72: ...pressing the VOLTAGE and CURRENT adjustment con trols at the same time then see PAR 3 2 8 If the panel is locked and RS 232 or GPIB access is not possible turn the power supply off set ANALOG I O DIP...

Page 73: ...latile memory either when EXIT is executed local mode PAR 4 3 4 or SAVE Calibration is executed remote mode PAR 4 4 4 Calibration of the KLP is performed in one of two ways locally using the front pan...

Page 74: ...GE or CURRENT control in either direction until status display reads CAL 7 Tap DC OUTPUT switch to enter calibration mode The status display shows PASS Enter the password see PAR 3 2 5 and tap DC OUTP...

Page 75: ...p the DC OUTPUT switch to accept the value The status display reads CMAX Rotate the CURRENT control to increase or decrease the DVM reading until the DVM reading corre sponds as close as possible to m...

Page 76: ...nt to enter the current date Rotate either the VOLTAGE or CURRENT controls to change the digit then tap either control to move to the next digit NOTE For purposes of traceability the user is strongly...

Page 77: ...ected across the sense terminals of the shunt resistor 1 Turn off power supply and disconnect load from output bus bar at the rear of the unit 2 Verify that power supply is configured for local error...

Page 78: ...PAR 4 4 4 to exit calibration 4 4 1 VOLTAGE CALIBRATION VXI plug play DRIVER DEMO During voltage calibration the voltage voltage readback and remote analog voltage readback are calibrated 1 Click the...

Page 79: ...ent readback are cali brated 1 Click the CURRENT button on the Calibration Window to start current calibration the Cur rent Calibration Window Figure 4 3 appears and the status display on the VXI main...

Page 80: ...to establish full scale output current and voltage are defined The factory default calibration of 10V programs full scale output however the following procedure allows any d c value up to 10V to progr...

Page 81: ...aximum rated voltage and current is set to the value corresponding to 1200 Watts If a current calibration is completed but no Voltage calibration the virtual model cur rent will be set for maximum rat...

Page 82: ...e moved to Previous and the new calibration values are stored in Working It is possible to replace the Working calibration values with either the Previous or the Factory calibration values contact Kep...

Page 83: ...oad see PAR 3 2 3 How do I establish the operating mode Operating mode is determined automatically by the programmed voltage Vp and current Ip settings and the load If the load impedance is greater th...

Page 84: ...tual model limits Programming inputs are only active when analog port is enabled See PAR 3 4 2 for details How do I Read back operating voltage and current Analog signals proportional to actual output...

Reviews: