3-32
BOP HIPWR 111315
2. Assign a name to the waveform: highlight Name, press
!
to modify, change the name
using the
Y
or
U
keys or
ADJUST
or the multiple presses of the numeric keys (see PAR.
3.2.4.1 for details). Press
$
to save or
%
to abort.the change and exit.
TABLE 3-11. WAVEFORM SEGMENT DETAILS MENU
SETTING
CHOICES
FUNCTION
Type
Square
Pos Ramp
Neg Ram
Triangle
Sine
Level
Trigger
Square - Square wave, bipolar, starts with positive excursion (see Note 1).
Pos Ramp - Increasing ramp, bipolar (see Note 1).
Neg Ramp - Decreasing ramp, bipolar (see Note 1).
Triangle - bipolar, starts with positive excursion, start/stop angle may be user controlled
(see Note 1).
Sine - - bipolar, starts with positive excursion, start/stop angle may be user controlled
(see Note 1).
Level - value determined by offset., duration determined by Period.
Trigger - Determines whether high level, low level or high-to-low edge is required at
trigger port to allow waveform to continue. NOTE: Offset does not affect trigger, 0V off-
set is recommended.
Frequency (Hz) or
Period (Sec) for Level
only
xxxx.xxx (Hz) or
xx.xxxx (Seconds)
(see Note 3)
Square wave: 0.02 to 1000 Hz
Ramp: 0.02 to 532 Hz
Sine or Triangle:0.001 to 443 Hz
Level: Period in Seconds. 0.005 to 5.0000 Seconds
Amplitude (P to P)
xxx.xxx
(Volts or Amperes)
Peak to peak amplitude of segment. Volts or Amperes determined by mode selected
for waveform. Value within model rating acceptable (e.g., for BOP 100-10MG, maxi-
mum amplitude is 200V). Values that exceed the maximum or minimum software limits
(see PAR. 3.3.4.2) are neither accepted nor executed.
Offset
xxx.xxx
(Volts or Amperes)
DC level on which the waveform (centered) will ride or Amplitude for Level segments.
Volts or Amperes determined by mode selected for waveform. E.g. for BOP 100-10MG,
to specify a positive ramp that goes from +30V to +80V, set amplitude to +50V and off-
set to +55V. Since p-p amplitude of segment is centered on offset, it is possible for the
waveform to exceed the model ratings (e.g., for the BHK 100-10MG, if the amplitude is
set to 150V and the offset is -50V, the negative portion of the segment will exceed the -
100V maximum of the model). In this case the output is clamped to the max/min soft-
ware limit in effect (see PAR. 3.3.4.1). Values that exceed the maximum or minimum
software limits (see PAR. 3.3.4.2) are neither accepted nor executed.
Start Angle
(Sine or Triangle only)
xxx x
(degrees)
The point at which the sine or triangle waveform segment starts. Values from 0.0° to
360.0° are acceptable. Default is 0.0°.
Stop Angle
(Sine or Triangle only)
xxx x
(degrees)
The point at which the sine or triangle waveform segment stops. Values from 0.0° to
360.0° acceptable. Default is 0.0° (functions the same as 360°).
Initial/ Repeat
Repeating
Initial
Repeating - Segment is executed for each cycle specified by the Count (see Note 2).
Initial - Segment is executed only on the first count and skipped for all subsequent
counts. Useful for preconditioning applications (see Note 2).
All segments except the last may be set to Initial or Repeating. The last segment is
always Repeating
NOTES
1. All waveforms except Level are centered on the offset. E.g. for a BOP 100-10MG, a positive ramp of 100V p-p amplitude, 10V
offset, will start at -40V and increase to +60V over the time period (T) determined by the frequency (F): T = 1/F.
2. When starting a new waveform, the first segment is always a repeating segment. One or more initial segments can be inserted
before the first repeating segment. Segments inserted before an initial segment are automatically set to initial; segments
inserted after a repeating segment are automatically set to repeating. It is recommended that the repeating segments be created
first. E.g. For a waveform with five segments, 1 and 2 designated Initial and 3, 4, and 5 designated Repeating, If the count was
set to 10, the first waveform would be comprised of segments 1 through 5, but the next nine waveforms would only have seg-
ments 3, 4, and 5. This feature can be useful for establishing an initial delay or preconditioning the load prior to running the
repeating waveform.
3. If the internal memory space used to process a waveform segment is exceeded, *** is displayed to the left of the segment(s) in
error. This can be corrected by 1) increasing the frequency (the lower the frequency, the more memory is used), 2) reducing the
number of segments.
Summary of Contents for BOP-MG 1KW
Page 2: ......
Page 10: ......
Page 20: ...x BOP 1K 111315 FIGURE 1 1 HIGH POWER BOP SERIES POWER SUPPLY...
Page 39: ...BOP HIPWR 111315 1 19 1 20 Blank FIGURE 1 3 BOP OUTPUT CHARACTERISTICS...
Page 40: ......
Page 56: ...2 16 BOP HIPWR 111315 FIGURE 2 5 PARALLEL CONFIGURATION LOCAL SENSING TYPICAL...
Page 57: ...BOP HIPWR 111315 2 17 FIGURE 2 6 PARALLEL CONFIGURATION REMOTE SENSING TYPICAL...
Page 68: ......