Operating instructions CellaCast PX 8x
19 Trouble shooting
19.1 No measured values are determined
The CellaCast measuring system checks the validity of the measured values for
each measurement and calculates the signal strength� The value for the signal
strength must be > 5%� The signal strength is only displayed during casting�
Parameter
Calculated signal strength
Q.Si6
In [%]
Checking the signal strength
► Press [Mode]
> The display shows [Ao]
► Press ▼ until [
Qu
] is displayed�
► Press [Mode]
> Im Display wird [
Q.EPS
] angezeigt
► Press ▼ until [
Q.Si6
] is displayed�
► Press [Mode]
> The display shows the currently calculated signal strength�
If the displayed signal strength is < 5 %, the following checks must be carried out:
Check lens or protective glass for dirt�
Check the alignment and focusing�
• The pyrometer must be focused to the measuring distance�
• The measuring field must be filled to 30 %�
• The pyrometer must have a clear view of the pouring stream�