18
Operating instructions CellaCast PX 8x
► In the functional menu press ▲ or ▼ to change to the parameter [
End
], then
press [Mode]�
If the time delay
T.DEL
must also be adjusted, proceed as described above� In
this case the parameter
T.DEL
must be selected and changed accordingly�
9.2 Setting the sampling time at the PX 80/81
The measuring time is set to 15 s in the delivery state� After the measuring time
has expired, the measured temperature is shown on the display or the analogue
output� Afterwards, the measuring time restarts automatically after the dead time
has elapsed�
Two-colour channel [
Qu
] → ATD function [
Q.Atd
]
► Press [Mode] to access the main menu
> The display shows [
Ao
]
► Press the ▼ until [
Qu
] is displayed�
► Press [Mode]
> The display shows [
Q.EPS
]
► Press ▼ until [
Q.ATD
] is displayed�
►Press [Mode]
> The display shows [
t.dEl
]
► Press ▼ until [
t.Act
] is displayed�
► Press [Mode]
> The current sampling time is displayed�
► Press ▲ or ▼ for 2 seconds
> The display flashes three times�
► Press ▲ or ▼ to set the desired sampling time.
► Press [Mode] to confirm the parameter
Exit operation
Parameter
Function
Default
T.DEL
Time delay [s]
1 s
T.ACT
Sampling [s]
15 s