9.1 Setting the sampling time at the PX 83
On delivery the pyrometer is parameterized on a discontinuous pouring stream� If
the pouring times are < 5 seconds the measuring time needs to be adapted� The
time delay (time after detection of the pouring stream up to the start of the meas-
urement) and the measurement time must be less than the minimum pouring time�
Two-colour channel [
Qu
] → ATD function [
Q.Atd
]
► Press [Mode] to access the main menu
> The display shows [
Ao
]
► Press ▼ until [
Qu
] is displayed�
► Press [Mode]
> The display shows [
Q.EPS
]
► Press ▼ until [
Q.ATD
] is displayed�
► Press [Mode]
> The display shows [
t.dEl
]
► Press ▼ until [
t.Act
] is displayed�
► Press [Mode]
> The current sampling time or
Auto
is displayed�
► Press ▲ or ▼ for 2 seconds
> The display flashes three times�
► Press ▲ or ▼ to set the desired sampling time.
► Press [Mode] to confirm the parameter
Exit operation
► Wait 30 seconds
or
► Press ▲ or ▼ to change to the paramete [
End
]� Then press [Mode] to
change to the functional menu�
Parameter
Function
Default
T.DEL
Time delay [s]
1 s
T.ACT
Sampling time [s]
Auto