16
Operating instructions CellaCast PX 8x
9 Parameters to be set for commissioning
On delivery the pyrometer is parameterized for the measurement on liquid metals�
The start of a measuring cycle is automatically detected and depends on limit 1,
limit 2 and the dead time� Before starting the measurement, Limit 1 must have
fallen below at least once� Limit 2 must be exceeded continuously for the duration
of the dead time (T�DEL)� If necessary, the limits are to be adapted�
Two-colour channel [
Qu
] → ATD function [
Q.Atd
]
1: Measuring object in front of the pyrometer
2: Limit 2 [
LI. 2
]
3: Limit 1 [
LI. 1
]
4: Time delay [
T.DEL
]
5: Sampling time [
T.ACT
]
Parameter
Function
Default
(
LI. 1
)
Limit 1
1100 °C
(
LI. 2
)
Limit 2
1200 °C