23
EN
Operating instructions CellaCast PX 8x
12.1 Processing of measuring signals
The pyrometer disposes of an IO-Link interface�
The generated three output signals correspond to the parameterization:
•
Out 1: Switching output/ IO-Link
■ Switch signal: threshold value temperature / status signal
•
Out 2: Switching output
■ Switch signal: threshold for temperature / status signal
•
Analogue output: 0/4 -20 mA
■ Analogue output for temperature
12.2 Analogue output
The pyrometer converts the measuring signal into a temperature proportional
analogue signal of 0/4 – 20 mA� The maximum load is 500 Ohm�
[
Ao.FN
] Changeover 0 -20 mA or 4 -20 mA
[
Ao.SP
] defines at which measured value the output signal is 0/4 mA�
[
Ao.EP
] defines at which measured value the output signal is 20 mA�
20.5
20
4
3.8
I [mA]
T [°C/°F]
1
2
20.5
20
4
3.8
I [mA]
T [°C/°F]
1
2
3
4
Maximum measuring range
Measuring range scaled
1 initial value of the measuring range
2 final value of the measuring range
3 Analogue start point
4 Analogue end point
Summary of Contents for CellaCast PX 80