Series 2200 Programmable DC Power Supplies Reference Manual
Appendix B: Calibration
2200S-901-01 Rev. C / March 2016
B-31
Instrument
Minimum test voltage
2200 - 20 - 5
18.0 V
2200 - 30 - 5
27.0 V
2200 - 32 - 3
28.8 V
2200 - 60 - 2
54.0 V
2200 - 72 - 1
64.8 V
9.
Enter the voltmeter reading into the DC current load regulation table for checking current load
regulation at the maximum voltage for your product.
10. Calculate the two values: REF – Min and Max – REF. Enter the values into the appropriate
columns in the DC current load regulation table.
11. Power off the DUT and test equipment.