Series 2200 Programmable DC Power Supplies Reference Manual
Section 2: General operation
2200S-901-01 Rev. C / March 2016
2-31
Step number
Voltage level (V)
Current level (A)
Duration (sec)
1
2
5
0.001
2
5
5
0.001
3
1
1
1.000
4
7
3
0.001
5
3
1
0.001
Keystrokes
Use the following procedures to configure the list parameters,create the list steps, save the list, run,
and stop running the list. In these procedures, you can use the arrow keys, numeric keypad, or the
navigation wheel
to select or enter numbers.
To configure the list parameters:
1. Press
Shift
and
1
to display the configuration menu.
2. Select
Edit List
. Press
Enter
. The display shows
Recall 1
.
3. Indicate that you want to save the list you are creating in location 3 by selecting
3
and pressing
Enter
. The display shows
Continuous
.
4. Indicate that you want the continuous mode by selecting
Continuous
and pressing
Enter
. The
display shows
Repeat 1
.
5. Indicate that you want to repeat the sequence twice by selecting
2 and p
ress
Enter
. The
display shows
List Steps 1.
6. Indicate that there will be
5
steps in the list by selecting 5 and pressing
Enter
. The display shows
S 001= 0.000 V
.
This completes entering the list parameters. You are ready to configure the first step.
To configure the first step in the list:
1. Input
S 001=
2.000
V
.
Press
Enter
. The display shows
S 001= 0.0000A
.
2. Input
S 001=
5.000
A
as the current limit.
Press Enter
. The display shows
S 001= 1.001S
.
3. Input
S 001=
0.001
S
as the duration of the step. Press
Enter
. The display shows
S 002=
0.000V
.
This completes the configuration of the first step. You are ready to configure the second step.
To configure the second step in the list:
1. Input
S 002=
5.000
V
. Press
Enter
. The display shows
S 002= 0.0000A
.
2. Input
S 002=
5.000
A
as the current limit. Press
Enter
. The display shows
S 002= 1.0000A
.
3. Input
S 002=
0.001
S
as the duration of the step. The display shows
S 002= 0.001S
. Press
Enter
. The display shows
S 003= 0.000V
.
This completes the configuration of the second step. You are ready to configure the third step.