Series 2200 Programmable DC Power Supplies Reference Manual
Section 4: Command reference
2200S-901-01 Rev. C / March 2016
4-19
[SOURce:]OUTPut:PON[:STATe]
This command configures the power supply to power up with its
output turned off, or to return the output to the state it was in when
it powered down.
Group
Source
Syntax
[SOURce:]OUTPut:PON[:STATe] {RST|RCL0}
[SOURce:]OUTPut:PON[:STATe]?
Arguments
RST
sets the power supply to power-up with output off.
RCL0
sets the power supply to power-up with the output in the last
state before power was removed.
Returns
RST|RCL0
Examples
OUTPUT:PON RST
OUTPUT:PON?
might return RCL0, which would indicate that the
instrument will return to the present output state if the power is
cycled.
[SOURce:]OUTPut:PROTection:CLEar (no query form)
This command clears a trip condition caused by over voltage (OV),
over temperature (OT), or remote inhibit (RI).
Group
Source
Syntax
[SOURce:]OUTPut:PROTection:CLEar
Related
Commands
[SOURce:]VOLTage:PROTection[:LEVel]
(on page 4-24)
Examples
OUTP:PROT:CLE