2-9
Notes:
If the output clock counter value is 0, the output clock pulse is
generated simultaneously with the update of the analog output channels
(no delay).
The DDA-08/16 can write analog output samples
after
the hardware
detects a digital trigger condition. Use post-triggering in DriverLINX
when you want to synchronize the start of data acquisition with an
external signal. Digital Start Events contain
mask, pattern,
and
match
fields. The mask is logically ANDed with the digital input data on the
Logical Channel and then compared with the
pattern
for a
match/mismatch.
●
Specify the
Channel
as
0
.
●
Specify the
Mask
and
Pattern
properties as
1
to specify the bit
position of the 1-bit trigger input.
●
Specify the
Match
property as
Not equals
to trigger on the edge of
the trigger input.
●
Specify the
Delay
property as any number of samples from
0 to 2
32
- 1.
●
Connect the signal to the TRIGGER IN line.
Refer to
Using DriverLINX with Your Hardware, Keithley DDA-08/16
manual that accompanies your DriverLINX software.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com