7-34
Scanning
Model 2750 Multimeter/Switch System User’s Manual
Operation
A simplified model of external trigger scan operation is shown in
, while the
procedure steps and programming commands are listed in
.
As shown in the operation model, when the scan is enabled, channel 101 closes and the
Model 2750 waits for an external trigger. When the trigger is received, channels 101 and
102 are measured. Operation then returns to the control source where it waits for another
trigger.
NOTE
After the scan is enabled (
, step 5), the TRIG key can be used to trigger
the scan.
Figure 7-4
External trigger scan example
2
Measurements
?
Measure
Open Last Chan
Close Next Chan
No
Yes
Wait For Trigger
Close Chan 101
Enable Scan
2750-900-01.book Page 34 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...