Triggering
8-13
What happens next depends on the state of initiation. If continuous initiation is already
enabled, the instrument will leave the idle state. SYSTem:PRESet enables continuous ini-
tiation. Therefore, operation will immediately leave the idle state when it is sent. The
*RCL command will do the same if INITiation:CONTinuous ON is a user saved default.
*RST disables continuous initiation. Therefore, the instrument will remain in the idle
state.
Either of the following two initiate commands will take the instrument out of the idle state:
•
INITiate
•
INITiate:CONTinuous ON
NOTE
While in remote, pressing the LOCAL key restores continuous front panel
operation.
2750-900-01.book Page 13 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...