Model 2750 Multimeter/Switch System User’s Manual
Range, Digits, Rate, Bandwidth, and Filter
4-9
The front panel RATE settings for all but the AC functions are explained as follow:
•
FAST sets integration time to 0.1 PLC. Use FAST if speed is of primary impor-
tance (at the expense of increased reading noise and fewer usable digits).
•
MEDium sets integration time to 1 PLC. Use MEDium when a compromise
between noise performance and speed is acceptable.
•
SLOW sets integration time to 5 PLC. SLOW provides better noise performance at
the expense of speed.
For the AC functions (ACV, ACV dB, and ACI), the RATE key sets integration time and
bandwidth. As listed in
, FAST sets NPLC to 1, while the MEDium and SLOW
NPLC settings are ignored (see
, for details).
From the front panel, setting the rate for one function affects all the other functions. For
example, if you set DCV for medium speed, the other functions will also set to medium
speed. For remote programming, each function can have its own unique rate setting (0.01
to 50 or 60 PLC).
NOTE
Rate cannot be set for continuity. It is fixed at 0.01 PLC.
Table 4-4
Rate and bandwidth settings
Function
Rate and bandwidth
Fast
Medium
Slow
DCV, DCI
ACV, ACI
Ω
2,
Ω
4
FREQ, PERIOD
Continuity
NPLC=0.1
NPLC=1, BW=300
NPLC=0.1
APER=0.01s
NPLC = 0.01
NPLC=1
NPLC=X, BW=30
NPLC=1
APER=0.1s
N/A
NPLC=5
NPLC=X, BW=3
NPLC=5
APER=1s
N/A
Notes: NPLC = number of power line cycles.
BW = lower limit of bandwidth (in Hz).
APER = aperture in seconds.
N/A = not available.
X = setting ignored.
2750-900-01.book Page 9 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...