Model 2750 Multimeter/Switch System User’s Manual
Status Structure
11-19
Event enable registers
shows, each status register set has an enable register. Each event register
bit is logically ANDed (&) to a corresponding enable bit of an enable register. Therefore,
when an event bit is set and the corresponding enable bit is set (as programmed by the
user), the output (summary) of the register will set to 1, which in turn sets the summary bit
of the Status Byte Register.
The commands to program and read the event enable registers are listed in
. For
details on programming and reading registers, see
“Programming enable registers,”
“Reading registers,” page 11-6
.
NOTE
The bits of any enable register can be reset to 0 by sending the 0 parameter
value with the appropriate enable command (i.e.
STATus:OPERation:ENABle 0).
Programming example — program and read register set
The following command sequence programs and reads the measurement register set. Keep
in mind that the Model 2750 has to be addressed to talk after a query command.
STAT:MEAS:ENAB 512
' Enable BFL (buffer full).
STAT:MEAS:COND?
' Read Measurement Condition Register.
STAT:MEAS?
' Read Measurement Event Register.
Table 11-5
Event enable registers commands
Command
Description
*ESE <NRf>
Program Standard Event Enable Register (0 to 255).
*ESE?
Read Standard Event Enable Register.
STATus:OPERation:ENABle <NRF>
Program Operation Event Enable Register (0 to 65535).
STATus:OPERation:ENABle?
Read enable register.
STATus:MEASurement:ENABle <NRf>
Program Measurement Event Enable Register (0 to 65535).
STATus:MEASurement:ENABle?
Read enable register.
STATus:QUEStionable:ENABle <NRf>
Program Questionable Event Enable Register (0 to 65535).
STATus:QUEStionable:ENABle?
Read enable register.
Note: Power-up and STATus:PRESet resets all bits of all enable registers to 0. *CLS has no effect.
2750-900-01.book Page 19 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...