7-22
Scanning
Model 2750 Multimeter/Switch System User’s Manual
NOTE
Channels for an advanced scan can be configured using different mX+B units
(i.e., ° and ¾), temperature sensors (I.e., 4-wire RTD and thermistor), and mea-
surement type (i.e., OCOMP ohms and DCV).
However, when readings are recalled from the buffer, the display may not indicate
the correct mX+B units symbol or annunciator for each channel. For example,
assume one channel used OCOMP ohms, while a second used ¾2. When the read-
ings are recalled, the OCOMP annunciator may remain on for both channels.
This display anomaly is due to memory limitations. Preserving the mX+B units
and annunciators for each channel would reduce the number of readings that
could be stored in the buffer.
2750-900-01.book Page 22 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...