Basic DMM Operation
Model 2750 Multimeter/Switch System User’s Manual
Current measurements (DCI and ACI)
The Model 2750 can make DCI measurements from 10nA to 3A and ACI measurements
from 1µA to 3A RMS.
NOTE
See the previous discussion about crest factor in
Connections
NOTE
When using the front panel inputs, the INPUTS switch must be in the “F” (out)
position. For switching modules, it must be in the “R” (in) position.
WARNING
To prevent electric shock, never make or break connections while
power is present in the test circuit.
Front panel inputs
When using the front panel input terminals, connect the test leads to the AMPS and
INPUT LO terminals as shown in
.
Figure 3-4
DCI and ACI connections using front panel inputs
Caution: Maximum Input = 3A DC or RMS
Model 2750
Front Panel
Current
Source
3A, 250V
AMPS
!
F
500V
PEAK
FRONT/REAR
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUT
350V
PEAK
1000V
PEAK
R
CAT I
2750-900-01.book Page 14 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...