B-6
Calibration Reference
Current measurement commands
:SENSe
(:CALibration:PROTected:SENSe[1])
(:CALibration:PROTected:SENSe2)
Purpose
To calibrate the active channel 1 or channel 2 current range.
Format
:cal:prot:sens1 <Cal_current>
:cal:prot:sens2 <Cal_current>
Parameters
<Cal_current> = Within ±10% of positive full-range value
0 ±0.1% of full-range value
Within ±10% of negative full-range value
Description
The :CAL:PROT:SENS1 and :CAL:PROT:SENS2 commands calibrate
the active range of the channel 1 or channel 2 current measurement
respectively. During the calibration process, this command is sent three
times for each range, once each with parameters of approximately
positive full range, 0, and negative full range. The appropriate
calibration current must be applied to the channel 1 or channel 2 input
terminals.
Example
:CAL:PROT:SENS1 1.9e-3
Calibrate channel 1 2mA range.
:DATA
(:CALibration:PROTected:SENSe[1]:DATA)
(:CALibration:PROTected:SENSe[1]:DATA?)
(:CALibration:PROTected:SENSe2:DATA)
(:CALibration:PROTected:SENSe2:DATA?)
Purpose
To set/query channel 1 or channel 2 measurement calibration constants.
Format
:cal:prot:sens1:data <Cal_constants>
:cal:prot:sens2:data <Cal_constants>
Parameters
<Cal_constants> = ASCII floating point format, delimited by commas
Query
:cal:prot:sens1:data?
:cal:prot:sens2:data?
Description
The :CAL:PROT:SENS1:DATA and :CAL:PROT:SENS2:DATA
commands allow you to send calibration constants directly to the unit,
while the :CAL:PROT:SENS1:DATA? and :CAL:PROT:SENS2:DATA?
queries request calibration constants. These commands allow you to
alter calibration directly without having to perform the entire calibration
procedure.
Example
:CAL:PROT:SENS1:DATA?
Query channel 1 constants.
Summary of Contents for 2500
Page 2: ...Contains Calibrating and Servicing Information Model 2500 Dual Photodiode Meter Service Manual...
Page 11: ......
Page 13: ......
Page 15: ......
Page 16: ...1 PerformanceVerification...
Page 28: ...2 Calibration...
Page 56: ...3 RoutineMaintenance...
Page 59: ...3 4 Routine Maintenance Model 2500 Service Manual...
Page 60: ...4 Troubleshooting...
Page 74: ...5 Disassembly...
Page 81: ...5 8 Disassembly Model 2500 Service Manual...
Page 82: ...6 ReplaceableParts...
Page 93: ...6 12 Replaceable Parts Model 2500 Service Manual...
Page 94: ...A Specifications...
Page 98: ...B CalibrationReference...
Page 108: ...C CalibrationProgram...
Page 117: ......
Page 119: ......
Page 120: ......
Page 121: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...