4-6
Troubleshooting
Analog circuits
shows a simplified block diagram of the analog circuits.
Current measurement circuits
Signal conditioning for the input current is provided by an I/V converter, which converts
the input signal current to voltage that can be used by the A/D converter. Current ranging
is provided by selecting various feedback resistors and switching in an X10 gain amplifier
at appropriate times. The multiplexer (not shown) switches among different signals during
the various phases of the measurement cycle, and passes the signal along to the A/D
converter.
A/D converter
Each Model 2500 channel uses a multi-slope charge balance A/D converter with a single-
slope run-down. The converter is controlled by a gate array. Commands are issued by the
MPU to the gate array, and the gate array IC sends A/D reading data back to the MPU for
calibration and processing.
Bias voltage circuits
Each Model 2500 voltage bias source is a digitally controlled source that can source up to
±100V @ 20mA. Digital control information from the MPU is converted into an equiva-
lent analog signal, which is amplified to provide the full ±100V, 20mA output capability.
Source ranging (10V or 100V) is performed by controlling the overall gain of the amplifier
stage. Compliance circuitry detects when the output current exceeds a fixed 20mA limit
and sends an over-compliance flag bit back to the MPU.
Summary of Contents for 2500
Page 2: ...Contains Calibrating and Servicing Information Model 2500 Dual Photodiode Meter Service Manual...
Page 11: ......
Page 13: ......
Page 15: ......
Page 16: ...1 PerformanceVerification...
Page 28: ...2 Calibration...
Page 56: ...3 RoutineMaintenance...
Page 59: ...3 4 Routine Maintenance Model 2500 Service Manual...
Page 60: ...4 Troubleshooting...
Page 74: ...5 Disassembly...
Page 81: ...5 8 Disassembly Model 2500 Service Manual...
Page 82: ...6 ReplaceableParts...
Page 93: ...6 12 Replaceable Parts Model 2500 Service Manual...
Page 94: ...A Specifications...
Page 98: ...B CalibrationReference...
Page 108: ...C CalibrationProgram...
Page 117: ......
Page 119: ......
Page 120: ......
Page 121: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...