![Keithley 2500 Service Manual Download Page 51](http://html1.mh-extra.com/html/keithley/2500/2500_service-manual_3998319051.webp)
2-24
Calibration
6.
Note and record the DMM reading, and send that value as the parameter for the fol-
lowing command:
:CAL:PROT:SOUR2 <DMM_Reading>
7.
Send the following command to output 0V:
:SOUR2:VOLT 0
8.
Note and record the DMM reading, and send that value as the parameter for the fol-
lowing command:
:CAL:PROT:SOUR2 <DMM_Reading>
9.
Send the following command to output -10V:
:SOUR2:VOLT -10
10.
Note and record the DMM reading, and send that value as the parameter for the fol-
lowing command:
:CAL:PROT:SOUR2 <DMM_Reading>
11.
Repeat steps 4 through 10 for the 100V range using
as a guide. Be sure
to:
•
Select the 100V range using the :SOUR2:VOLT:RANG 100 command.
•
Send the ±100V source values where appropriate.
12.
Send this command to turn off the channel 1 output:
:OUTP2 OFF
Table 2-9
Channel 2 remote voltage bias calibration summary
Voltage bias
range
Calibration
voltages
Calibration commands
10V
+10V
0V
-10V
:OUTP2 ON
:SOUR2:VOLT:RANG 10
:SOUR2:VOLT 10
:CAL:PROT:SOUR2 <DMM_Reading>
:SOUR2:VOLT 0
:CAL:PROT:SOUR2 <DMM_Reading>
:SOUR2:VOLT -10
:CAL:PROT:SOUR2 <DMM_Reading>
100V
+100V
0V
-100V
:SOUR2:VOLT:RANG 100
:SOUR2:VOLT 100
:CAL:PROT:SOUR2 <DMM_Reading>
:SOUR2:VOLT 0
:CAL:PROT:SOUR2 <DMM_Reading>
:SOUR2:VOLT -100
:CAL:PROT:SOUR2 <DMM_Reading>
:OUTP2 OFF
Summary of Contents for 2500
Page 2: ...Contains Calibrating and Servicing Information Model 2500 Dual Photodiode Meter Service Manual...
Page 11: ......
Page 13: ......
Page 15: ......
Page 16: ...1 PerformanceVerification...
Page 28: ...2 Calibration...
Page 56: ...3 RoutineMaintenance...
Page 59: ...3 4 Routine Maintenance Model 2500 Service Manual...
Page 60: ...4 Troubleshooting...
Page 74: ...5 Disassembly...
Page 81: ...5 8 Disassembly Model 2500 Service Manual...
Page 82: ...6 ReplaceableParts...
Page 93: ...6 12 Replaceable Parts Model 2500 Service Manual...
Page 94: ...A Specifications...
Page 98: ...B CalibrationReference...
Page 108: ...C CalibrationProgram...
Page 117: ......
Page 119: ......
Page 120: ......
Page 121: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...