B-8
Calibration Reference
:OFFSet
(:CALibration:PROTected:SOURce[1]:OFFSet)
(:CALibration:PROTected:SOURce2:OFFSet)
Purpose
To calibrate the channel 1 or channel 2 input voltage offset.
Format
:cal:prot:sour1:offs
:cal:prot:sour2:offs
Description
The :CAL:PROT:SOUR1:OFFS and :CAL:PROT:SOUR2:OFFS com-
mands calibrate the channel 1 or channel 2 input voltage offset respec-
tively. A triax shielding cap should be installed on the corresponding
INPUT jack during these calibration steps.
Example
:CAL:PROT:SOUR2:OFFS
Calibrate channel 2 input voltage offset.
Detecting calibration errors
If an error occurs during any calibration step, the Model 2500 will generate an appropriate
error message. Several methods to detect calibration errors are discussed below.
Reading the error queue
As with other Model 2500 errors, any calibration errors will be reported in the error queue.
(You can read the error queue by using the :SYST:ERR? query.)
Error summary
summarizes calibration errors.
Table B-2
Calibration errors
Error number Error message
+500
“Date of calibration not set”
+501
“Next date of calibration not set”
+502
“Calibration data invalid”
+509
“Not permitted with cal locked”
+510
“Not permitted with cal unlocked”
+520
“ gain data invalid”
+521
“ offset data invalid”
+522
“Source - gain data invalid”
+523
“Source - offset data invalid”
+524
“Source DAC Overflow”
+525
“Source DAC Underflow”
Summary of Contents for 2500
Page 2: ...Contains Calibrating and Servicing Information Model 2500 Dual Photodiode Meter Service Manual...
Page 11: ......
Page 13: ......
Page 15: ......
Page 16: ...1 PerformanceVerification...
Page 28: ...2 Calibration...
Page 56: ...3 RoutineMaintenance...
Page 59: ...3 4 Routine Maintenance Model 2500 Service Manual...
Page 60: ...4 Troubleshooting...
Page 74: ...5 Disassembly...
Page 81: ...5 8 Disassembly Model 2500 Service Manual...
Page 82: ...6 ReplaceableParts...
Page 93: ...6 12 Replaceable Parts Model 2500 Service Manual...
Page 94: ...A Specifications...
Page 98: ...B CalibrationReference...
Page 108: ...C CalibrationProgram...
Page 117: ......
Page 119: ......
Page 120: ......
Page 121: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...