A-4
Specifications
Accuracy calculations
The following information discusses how to calculate accuracy for both current measure-
ment and voltage bias functions.
Current measurement accuracy
Current measurement accuracy is calculated as follows:
Accuracy = ±(% of r offset)
As an example of how to calculate the actual reading limits, assume that you are measur-
ing 1mA on the 2mA range. You can compute the reading limit range from one-year mea-
surement specifications as follows:
Accuracy = ±(% of r offset)
= ±
[(0.1% ×
1mA) + 50nA]
= ±(1
µ
A + 50nA)
= ±1.05
µ
A
Thus, the actual reading range is 1mA ±1.05
µ
A or from 0.99895mA to 1.00105mA.
Voltage bias accuracy
Voltage bias accuracy is calculated similarly, except that voltage bias specifications are
used. As an example of how to calculate the actual source output limits, assume that you
are sourcing 5V on the 10V range. You can compute the output limits from voltage bias
one-year accuracy specifications as follows:
Accuracy = ±(0.05% of s 5mV offset)
= ±
[(0.05% × 5
V
) + 5
mV
)]
= ±(2.5mV + 5mV)
= ±7.5mV
In this case, the actual voltage output range is 5V ±7.5mV or from 4.9925V to 5.0075V.
Summary of Contents for 2500
Page 2: ...Contains Calibrating and Servicing Information Model 2500 Dual Photodiode Meter Service Manual...
Page 11: ......
Page 13: ......
Page 15: ......
Page 16: ...1 PerformanceVerification...
Page 28: ...2 Calibration...
Page 56: ...3 RoutineMaintenance...
Page 59: ...3 4 Routine Maintenance Model 2500 Service Manual...
Page 60: ...4 Troubleshooting...
Page 74: ...5 Disassembly...
Page 81: ...5 8 Disassembly Model 2500 Service Manual...
Page 82: ...6 ReplaceableParts...
Page 93: ...6 12 Replaceable Parts Model 2500 Service Manual...
Page 94: ...A Specifications...
Page 98: ...B CalibrationReference...
Page 108: ...C CalibrationProgram...
Page 117: ......
Page 119: ......
Page 120: ......
Page 121: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...