1-6
Performance Verification
Performing the verification test procedures
Test summary
•
Current measurement accuracy
•
Voltage bias accuracy
If the Model 2500 is not within specifications and not under warranty, see the calibration
procedures in Section 2 for information on calibrating the unit.
Test considerations
When performing the verification procedures:
•
Be sure to restore factory front panel defaults as previously outlined.
•
Make sure that the test equipment is fully warmed up and properly connected to the
correct Model 2500 terminals as required.
•
Allow signals to settle before making a measurement.
•
Do not connect test equipment to the Model 2500 through a scanner, multiplexer,
or other switching equipment.
WARNING
The maximum common-mode voltage (voltage between LO and
chassis ground) is 200V DC. Exceeding this value may cause a shock
hazard.
CAUTION
Maximum signal on OUTPUT connectors is 100V @ 20mA DC.
Maximum signal on INPUT connectors is 20mA. Exceeding these
values may result in damage to the instrument.
Current measurement accuracy
Follow the steps below to verify that Model 2500 current measurement accuracy is within
specified limits. This test involves applying currents from a calibrator and verifying that
Model 2500 current readings are within required limits.
1.
With the power off, connect the calibrator to the INPUT CHANNEL 1 triax jack,
as shown in
.
2.
Turn on the Model 2500 and calibrator, and allow them to warm up for at least one
hour.
3.
Restore front panel (BENCH) defaults as outlined in “Restoring factory defaults.”
Summary of Contents for 2500
Page 2: ...Contains Calibrating and Servicing Information Model 2500 Dual Photodiode Meter Service Manual...
Page 11: ......
Page 13: ......
Page 15: ......
Page 16: ...1 PerformanceVerification...
Page 28: ...2 Calibration...
Page 56: ...3 RoutineMaintenance...
Page 59: ...3 4 Routine Maintenance Model 2500 Service Manual...
Page 60: ...4 Troubleshooting...
Page 74: ...5 Disassembly...
Page 81: ...5 8 Disassembly Model 2500 Service Manual...
Page 82: ...6 ReplaceableParts...
Page 93: ...6 12 Replaceable Parts Model 2500 Service Manual...
Page 94: ...A Specifications...
Page 98: ...B CalibrationReference...
Page 108: ...C CalibrationProgram...
Page 117: ......
Page 119: ......
Page 120: ......
Page 121: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...