3-2
Routine Maintenance
Introduction
The information in this section deals with routine type maintenance that can be performed
by the operator and includes information on line voltage selection and line fuse
replacement.
Line voltage selection
WARNING
Disconnect the line cord at the rear panel and remove all test leads
connected to the instrument before changing the line voltage.
The Model 2500 operates from a line voltage in the range of 100V / 120V / 220V / 240V,
±10% at a frequency of 50 or 60Hz. Before plugging in the unit, make sure the line voltage
setting seen through the small window in the power module (
) is correct for the
line voltage in your area.
WARNING
Operating the instrument on an incorrect line voltage may cause dam-
age, possibly voiding the warranty.
If the voltage setting is not correct, change it as outlined below.
1.
Pry the small tab that secures the fuse holder to the power module, then remove the
fuse holder.
2.
Pull out and rotate the fuse carrier to the correct line voltage setting, then install it
in the fuse holder. Make sure the line voltage shown in the window is correct. Also
make sure the fuse rating for the expected line voltage is correct. (See “Line fuse
replacement.”)
3.
Reinstall the fuse holder in the power module, making sure it is seated fully.
Line fuse replacement
WARNING
Disconnect the line cord at the rear panel and remove all test leads
connected to the instrument before replacing the line fuse.
The power line fuse is accessible from the rear panel and is integral with the AC power
module (
Summary of Contents for 2500
Page 2: ...Contains Calibrating and Servicing Information Model 2500 Dual Photodiode Meter Service Manual...
Page 11: ......
Page 13: ......
Page 15: ......
Page 16: ...1 PerformanceVerification...
Page 28: ...2 Calibration...
Page 56: ...3 RoutineMaintenance...
Page 59: ...3 4 Routine Maintenance Model 2500 Service Manual...
Page 60: ...4 Troubleshooting...
Page 74: ...5 Disassembly...
Page 81: ...5 8 Disassembly Model 2500 Service Manual...
Page 82: ...6 ReplaceableParts...
Page 93: ...6 12 Replaceable Parts Model 2500 Service Manual...
Page 94: ...A Specifications...
Page 98: ...B CalibrationReference...
Page 108: ...C CalibrationProgram...
Page 117: ......
Page 119: ......
Page 120: ......
Page 121: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...