TYPE:INT
MIN CHAN: 1
MAX CHAN: 10
TIMER?
TIMER? ON
00H:00M:05.000S
RDG CNT: 0010
STEP
10 channel closures
at 5-second intervals
10 output triggers
SCAN
10 channel closures
1 output trigger
RECALL
10 readings
RDG CNT: 0010
DELAY: MAN
00H:00M:05.000S
STEP
10 channel closures
at 5-second intervals
10 output triggers
SCAN
10 channel closures
at 5-second intervals
1 output trigger
RECALLL
10 readings
OFF
ON
Note: "Factory setup" on the
Model 2000 is assumed.
SHIFT-CONFIG
Figure 3-15
Internal scanning
example with tim-
er and delay op-
tions
Measurement Options
3-29