Connections
Assuming factory default conditions, the basic procedure is as follows:
1.
Connect test leads to the INPUT HI and LO terminals of the Model 2000. Either the
front or rear inputs can be used; place the INPUTS button in the appropriate position.
2.
Select the FREQ or PERIOD function.
3.
Connect test leads to the source as shown in Figure 2-7.
CAUTION
Do not exceed 1000V peak between INPUT HI and INPUT LO or instru-
ment damage may occur.
4.
Take a reading from the display.
See Section 3—Measurement Options for information that explains the configuration options
for frequency and period measurements.
2001 MULTIMETER
Model 2000
Caution: Maximum Input = 1000V peak, 8 x 10
7
V•Hz
AC Voltage
Source
Input Impedance = 1M
Ω
in parallel with <100pF
SHIFT
CH1
REM
TALK
LSTN
SRQ
STAT
REL
FILT
4W
BUFFER
MATH
REAR
SCAN
TIMER
STEP
CH2
CH3
CH4
CH5
CH6
CH7
CH
8
CH9
CH1
0
HOLD TRIG
FAST
MED
SLOW
AUTO
ERR
Basic Measurements
2-27
Figure 2-7
Frequency and
period measure-
ments
Summary of Contents for 2000
Page 1: ...Model 2000Multimeter User s Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 13: ...1 General Information...
Page 19: ...2 Basic Measurements...
Page 54: ...2 36 Basic Measurements...
Page 55: ...3 Measurement Options...
Page 87: ...4 Remote Operation...
Page 139: ...5 SCPI CommandReference...
Page 215: ...A Specifications...
Page 226: ...A 12 Specifications...
Page 227: ...B Statusand ErrorMessages...
Page 233: ...C Example Programs...
Page 246: ...C 14 Example Programs...
Page 247: ...D Models196 199and 8840A 8842ACommands...
Page 255: ...E IEEE 488 BusOverview...