[SENSe[1]] subsystem
The Sense 1 Subsystem is used to configure and control the measurement
functions of the Model 2000. A function does not have to be selected before
you program its various configurations. A function can be selected any time
after it has been programmed. Whenever a programmed function is selected,
it assumes the programmed states.
:FUNCtion Command
:FUNCtion <name>
[:SENSe[1]]:FUNCtion <name>
Select measurement function.
Parameters
<name> = ‘CURRent:AC’
‘Select AC Current
‘CURRent[:DC]’
‘Select DC Current
‘VOLTage:AC’
‘Select AC Voltage
‘VOLTage[:DC]’
‘Select DC Voltage
‘RESistance’
‘Select 2-wire Resistance
‘FRESistance’
‘Select 4-wire Resistance
‘PERiod’
‘Select Period
‘FREQuency’
‘Select Frequency
‘TEMPerature’
‘Select Temperature
‘DIODe’
‘Select Diode Testing
‘CONTinuity’
‘Select Continuity Testing
Query
:FUNCtion?
Query currently programmed function.
Description
The :FUNCtion command is used to select the measurement function of the
instrument. Note that parameter names are enclosed in single quotes (’).
However, double quotes (“) can instead be used. For example:
:func ‘volt’ = :func “volt”
Each measurement function “remembers” its own unique setup configura-
tion, such as range, speed, resolution, filter and rel. This eliminates the need
to re-program setup conditions every time you switch from one function to
another.
SCPI Command Reference
5-37
Summary of Contents for 2000
Page 1: ...Model 2000Multimeter User s Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 13: ...1 General Information...
Page 19: ...2 Basic Measurements...
Page 54: ...2 36 Basic Measurements...
Page 55: ...3 Measurement Options...
Page 87: ...4 Remote Operation...
Page 139: ...5 SCPI CommandReference...
Page 215: ...A Specifications...
Page 226: ...A 12 Specifications...
Page 227: ...B Statusand ErrorMessages...
Page 233: ...C Example Programs...
Page 246: ...C 14 Example Programs...
Page 247: ...D Models196 199and 8840A 8842ACommands...
Page 255: ...E IEEE 488 BusOverview...