SCPI Command Reference
5-59
:QUEue commands
[:NEXT]?
:STATus:QUEue[:NEXT]?
Read Error Queue
Description
As error and status messages occur, they are placed into the Error Queue.
This query command is used to read those messages.
The Error Queue is a first-in, first-out (FIFO) register. Every time you read
the queue, the “oldest” message is read and that message is then removed
from the queue. The queue will hold up to 10 messages. If the queue be-
comes full, the message “350, ‘Queue Overflow” will occupy the last mem-
ory location in the register. On power-up, the Error Queue is empty. When
the Error Queue is empty, the message “0, ‘No error” is placed in the Error
Queue.
The messages in the queue are preceded by a number. Negative (-) numbers
are used for SCPI defined messages, and positive (+) numbers are used for
Keithley defined messages. The messages are listed in Appendix B.
After this command is sent and the Model 2000 is addressed to talk, the “old-
est” message in the queue is sent to the computer.
NOTE
The :STATus:QUEue[:NEXT]? query command performs the
same function as the :SYSTem:ERRor? query command (see Sys-
tem subsystem).
:CLEar
:STATus:QUEue:CLEar
Clear Error Queue
Description
This action command is used to clear the Error Queue of messages.
Summary of Contents for 2000
Page 1: ...Model 2000Multimeter User s Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 13: ...1 General Information...
Page 19: ...2 Basic Measurements...
Page 54: ...2 36 Basic Measurements...
Page 55: ...3 Measurement Options...
Page 87: ...4 Remote Operation...
Page 139: ...5 SCPI CommandReference...
Page 215: ...A Specifications...
Page 226: ...A 12 Specifications...
Page 227: ...B Statusand ErrorMessages...
Page 233: ...C Example Programs...
Page 246: ...C 14 Example Programs...
Page 247: ...D Models196 199and 8840A 8842ACommands...
Page 255: ...E IEEE 488 BusOverview...