Trigger mode
T0 (default)
T1
T2
T3
T4
Trigger
Rear panel
Auto
mode
trigger
delay
Internal
Disabled
—
External
Enabled
On
External
Disabled
On
External
Enabled
Off
External
Disabled
Off
Note: Delay is enabled by entering EXT TRIG mode while in
local.
Offset (Rel)
B0
B1
Offset off (default)
Offset on
Display
D0
D1
Normal display (default)
Blank display
Suffix
Y0
Y1
Disable output suffix (default)
Enable output suffix
Terminator
W0
W1
W2
W3
W4
W5
W6
W7
Enable CR LF EOI (default)
Enable CR LF only
Enable CR EOI only
Enable CR only
Enable LF EOI only
Enable LF only
Enable EOI only
Disable all output terminators
Clear
*
X0
Device clear (resets 8842A to default conditions)
Clear error register
Single-trigger
?
Trigger measurement
GET
G0
G1
G2
G3
G4
G5
G6
G7
G8
Get instrument configuration (function, range, speed,
trigger)
Get SRQ mask
Get calibration input prompt (generates error message
51)
Get user-defined message (replies with 16 space charac-
ters)
Get calibration status (replies with “1000”)
Get IAB status (input F/R, autorange on/off, offset on/
off)
Get YW status (suffix enabled/disabled, terminator
selection)
Get error status
Get instrument ID
(replies with “FLUKE,8842A,0,EMUL”)
Table D-2 (cont.)
Models 8840A/8842A device-dependent command summary
Mode
Command
Description
Models 196/199 and 8840A/8842A Commands
D-7
Summary of Contents for 2000
Page 1: ...Model 2000Multimeter User s Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 13: ...1 General Information...
Page 19: ...2 Basic Measurements...
Page 54: ...2 36 Basic Measurements...
Page 55: ...3 Measurement Options...
Page 87: ...4 Remote Operation...
Page 139: ...5 SCPI CommandReference...
Page 215: ...A Specifications...
Page 226: ...A 12 Specifications...
Page 227: ...B Statusand ErrorMessages...
Page 233: ...C Example Programs...
Page 246: ...C 14 Example Programs...
Page 247: ...D Models196 199and 8840A 8842ACommands...
Page 255: ...E IEEE 488 BusOverview...