:AVERage commands
The :AVERage commands are used to configure and control the filter. The
Filter is explained in Section 3.
:STATe <b>
[:SENSe[1]]:CURRent:AC:AVERage:STATe <b>
Control filter for ACI
[:SENSe[1]]:CURRent[:DC]:AVERage:STATe <b>
Control filter for DCI
[:SENSe[1]]:VOLTage:AC:AVERage:STATe <b>
Control filter for ACV
[:SENSe[1]]:VOLTage[:DC]:AVERage:STATe <b>
Control filter for DCV
[:SENSe[1]]:RESistance:AVERage:STATe <b>
Control filter for
Ω
2
[:SENSe[1]]:FRESistance:AVERage:STATe <b>
Control filter for
Ω
4
[:SENSe[1]]:TEMPerature:AVERage:STATe <b>
Control filter for TEMP
Parameters
<b> = 0 or OFF Disable the digital filter
1 or ON
Enable the digital filter
Query
:STATe?
Query state of digital filter
Description
These commands are used to enable or disable the digital averaging filter for
the specified function. When enabled, readings will be filtered according to
how the filter is configured.
:TCONtrol <name>
:SENSe[1]]:CURRent:AC:AVERage:TCONtrol <name>
Select filter type for ACI
[:SENSe[1]]:CURRent[:DC]:AVERage:TCONtrol <name>
Select filter type for DCI
[:SENSe[1]]:VOLTage:AC:AVERage:TCONtrol <name>
Select filter type for ACV
[:SENSe[1]]:VOLTage[:DC]:AVERage:TCONtrol <name>
Select filter type for DCV
[:SENSe[1]]:RESistance:AVERage:TCONtrol <name>
Select filter type for
Ω
2
[:SENSe[1]]:FRESistance:AVERage:TCONtrol <name>
Select filter type for
Ω
4
[:SENSe[1]]:TEMPerature:AVERage:TCONtrol <name>
Select filter type for TEMP
Parameters
<name> =
REPeat
Select repeating filter
MOVing
Select moving filter
Query
:TCONtrol? Query filter type
Description
These commands are used to select the type of averaging filter (REPeat or
MOVing) for the specified function. These filter types are explained in Sec-
tion 3 (see “Filter Modes”).
The number of readings that are averaged by the filter is set with the :AVER-
age:COUNt command. The :AVERage:STATe command is used to enable or
disable the filter. Changing the filter type disables auto filter.
5-46
SCPI Command Reference
Summary of Contents for 2000
Page 1: ...Model 2000Multimeter User s Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 13: ...1 General Information...
Page 19: ...2 Basic Measurements...
Page 54: ...2 36 Basic Measurements...
Page 55: ...3 Measurement Options...
Page 87: ...4 Remote Operation...
Page 139: ...5 SCPI CommandReference...
Page 215: ...A Specifications...
Page 226: ...A 12 Specifications...
Page 227: ...B Statusand ErrorMessages...
Page 233: ...C Example Programs...
Page 246: ...C 14 Example Programs...
Page 247: ...D Models196 199and 8840A 8842ACommands...
Page 255: ...E IEEE 488 BusOverview...