Enable registers
As Figure 4-5 shows, each status register set has an enable register. An enable register is pro-
grammed by you and serves as a mask for the corresponding event register. An event bit is
masked when the corresponding bit in the enable register is cleared (0). When masked, a set bit
in an event register cannot set a bit in the Status Byte Register (1 AND 0 = 0).
To use the Status Byte Register to detect events (i.e., serial poll), you must unmask the events
by setting (1) the appropriate bits of the enable registers.
To program and query the Standard Event Status Register, use the *ESE and *ESE? Common
Commands respectively. All other enable registers are programmed and queried using the :EN-
ABle and :ENABLe? commands in the STATus Subsystem. See Section 5 for more information.
An enable register is not cleared when it is read. The following operations affect the enable
registers:
•
Cycling power - Clears all enable registers
•
:STATus:PREset clears the following enable registers:
Operation Event Enable Register
Questionable Event Enable Register
Measurement Event Enable Register
*ESE 0 - Clears the Standard Event Status Enable Register.
Remote Operation
4-21
Summary of Contents for 2000
Page 1: ...Model 2000Multimeter User s Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 13: ...1 General Information...
Page 19: ...2 Basic Measurements...
Page 54: ...2 36 Basic Measurements...
Page 55: ...3 Measurement Options...
Page 87: ...4 Remote Operation...
Page 139: ...5 SCPI CommandReference...
Page 215: ...A Specifications...
Page 226: ...A 12 Specifications...
Page 227: ...B Statusand ErrorMessages...
Page 233: ...C Example Programs...
Page 246: ...C 14 Example Programs...
Page 247: ...D Models196 199and 8840A 8842ACommands...
Page 255: ...E IEEE 488 BusOverview...