Configuration
The following information explains the various configuration options for temperature mea
surements. To select and configure the thermocouple measurement:
Press SHIFT then TCOUPL. Three choices are available using the
▲
and
▼
keys:
•
UNITS — C, K, F (Centigrade, Kelvin, Fahrenheit). This parameter selects the displayed
units for temperature measurements.
•
TYPE — J, K, T (thermocouple type).
•
JUNC — SIM, CH1 (simulated or referenced at Channel 1). Typically, a thermocouple
card uses a single reference junction. The Model 2000 can simulate a reference junction
temperature or use the reference junction on a switching card. Typical reference junction
temperatures are 0°C and 23°C.
A simulated reference temperature is the temperature of the junction where the thermocouple
voltage is sensed. It is room temperature if the thermocouple wire is terminated to banana jacks
and corrected directly to the multimeter. The accuracy of a temperature measurement depends
on the accuracy of the reference junction.
Basic Measurements
2-29
Summary of Contents for 2000
Page 1: ...Model 2000Multimeter User s Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 13: ...1 General Information...
Page 19: ...2 Basic Measurements...
Page 54: ...2 36 Basic Measurements...
Page 55: ...3 Measurement Options...
Page 87: ...4 Remote Operation...
Page 139: ...5 SCPI CommandReference...
Page 215: ...A Specifications...
Page 226: ...A 12 Specifications...
Page 227: ...B Statusand ErrorMessages...
Page 233: ...C Example Programs...
Page 246: ...C 14 Example Programs...
Page 247: ...D Models196 199and 8840A 8842ACommands...
Page 255: ...E IEEE 488 BusOverview...