:MUNits
:CALCulate [1]:KMATh:MUNits <name>
Specify units for mx+b
Parameter
<name> = 3 characters using ‘A’ through ‘Z’
Query
:MUNits?
Query units for mx+b
Description
This command is used to specify the units data element for the mx+b calcu-
lation. Use any three letters from ‘A’ through ‘Z’.
:PERCent <NRf>
:CALCulate [1]:KMATh:PERCent <NRf>
Specify target value for percent calculation
Parameter
<Nrf> = -1e8 to +1e8 Specify target value.
Query
:PERCENt?
Query percent target value
Description
This command is used to specify the target value for the percent calculation.
:ACQuire
:CALCulate [1]:KMATh:PERCent:ACQuire
Use input signal as target value
Description
This action command is used to acquire the present input signal reading and
use it as the target value for the Percent calculation.
:STATe <b>
:CALCulate [1]:STATe <b>
Control CALC1
Parameters
<b> =
0 or off
Disable CALC1 calculation
1 or on
Enable CALC1 calculation
Query
:STATe?
Query state (on or off) of CALC1.
Description
This command is used to enable or disable the CALC1 calculation. When
enabled, each instrument reading will reflect the selected calculation (see
:FORMat).
:DATA?
:CALCulate [1]:KMATh:DATA?
Read CALC1 result
Description
This query command is used to read the result of the CALC1 calculation. If
CALC1 is disabled or NONE is selected, the “raw” reading will be read.
SCPI Command Reference
5-21
Summary of Contents for 2000
Page 1: ...Model 2000Multimeter User s Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 13: ...1 General Information...
Page 19: ...2 Basic Measurements...
Page 54: ...2 36 Basic Measurements...
Page 55: ...3 Measurement Options...
Page 87: ...4 Remote Operation...
Page 139: ...5 SCPI CommandReference...
Page 215: ...A Specifications...
Page 226: ...A 12 Specifications...
Page 227: ...B Statusand ErrorMessages...
Page 233: ...C Example Programs...
Page 246: ...C 14 Example Programs...
Page 247: ...D Models196 199and 8840A 8842ACommands...
Page 255: ...E IEEE 488 BusOverview...