CAUTION

Risk of burn injury from hot handpiece head or hot handpiece lid.

Burn injuries in the mouth may be caused if the instrument overheats.

▶ Never touch soft tissue with the handpiece head or handpiece lid!

CAUTION

Risk due to incorrectly stored instrument.

Damage to clamping system from dropping the instrument.

▶ After treatment, place the instrument properly in the cradle without

the tool.

Safety

11

CAUTION

Instructions for the safe use of handpieces with electronic micromotors.

Electronic micromotors generate much more energy than conventional

pneumatic turbines and motors. Given the higher torque and speed, hand‐

pieces that are poorly serviced, damaged or used improperly can overheat

which can seriously burn the patient.

▶ Never contact soft tissue with the instrument head.

Note
For safety reasons, we recommend that the tool holder system be checked

annually after the warranty period expires.

Safety

12

The following individuals are authorized to repair and service KaVo prod‐

ucts:

▪ Technicians at KaVo branches throughout the world

▪ Technicians specially trained by KaVo

The following guidelines must be observed to ensure save use of the elec‐

trically driven handpieces:

▪ The service instructions for handpieces must be precisely following

when using KAVO spray or QUATTROcare care systems.

▪ Before each use, the handpiece must be checked for external damage.

▪ Before each use, perform a test run with the handpiece, and watch for

atypical heating and unusual noise and vibration.

▪ Immediately stop using handpieces that act unusual.

▪ Never press the pushbutton during operation. This also includes lifting

the cheek or tongue!

 

Safety

13

To ensure proper function, the medical device must be set up according to

the reprocessing methods described in the KaVo Instructions for Use, and

the care products and care systems described therein must be used. KaVo

recommends specifying a service interval at the dental office for a licensed

shop to clean, service and check the functioning of the medical device. This

service interval depends on the frequency of use and should be adjusted

accordingly.

Service may only be carried out by KaVo-trained repair shops using original

KaVo replacement parts.

 

Safety

14

Summary of Contents for INTRAmatic contra-angle 80 ES

Page 1: ...ger levels 8 2 2 Safety instructions 9 3 Product description 15 3 1 Purpose Proper use 16 3 2 Technical data 18 3 3 Transportation and storage conditions 19 Contents 1 4 4 First use 21 4 1 Check the a...

Page 2: ...Sterilisation 52 7 8 Storage 55 8 8 Tools and consumables 56 9 Terms and conditions of warranty 58 Contents 4 1 1 U User instructions Dear User Congratulations on purchasing this KaVo quality product...

Page 3: ...NING WARNING indicates a hazardous situation that can cause death or serious injury Safety 8 DANGER DANGER indicates a hazardous situation that can directly cause death or serious injury 2 2 S Safety...

Page 4: ...s at KaVo branches throughout the world Technicians specially trained by KaVo The following guidelines must be observed to ensure save use of the elec trically driven handpieces The service instructio...

Page 5: ...l device may only be used for the described application by a knowledgeable user The following must be observed the applicable health and safety regulations the applicable accident prevention regulatio...

Page 6: ...RH to 95 RH Product description 19 Air pressure 700 hPa to 1060 hPa 10 psi to 15 psi Protect from moisture Product description 20 4 4 F First use WARNING Hazard from nonsterile products Infection dan...

Page 7: ...eration 24 5 5 1 A Attaching the medical device Place the medical device on the LUX motor coupling and lock it into place Before each treatment pull on the medical device to see if it is securely seat...

Page 8: ...r or grinder is securely attached by pulling on it Operation 28 5 5 4 R Removing the milling tool or diamond grinder CAUTION Hazard from rotating tools Laceration Do not carelessly contact rotating to...

Page 9: ...1 T Troubleshooting Exchanging the O rings on the motor coupling CAUTION Hazard from improper care of the O rings Malfunctions or complete failure of the product Do not use Vaseline or other grease o...

Page 10: ...fectors in accordance with EN ISO 15883 that are operated with alkaline cleaning agents at a pH of max 10 e g Miele G 7781 G 7881 validation was performed with the VARIO TD programme neodisher medicle...

Page 11: ...ing agent and extends only to the compatibility of materials with respect to KaVo products For program settings as well as cleansers and disinfectants to be used please refer to the Instructions for U...

Page 12: ...Z neu tralisation agent and neodisher mielclear rinsing agent and extends only to the compatibility of materials with respect to KaVo products For program settings as well as cleansers and disinfecta...

Page 13: ...le tip in the opening and spray Carry out the servicing according to the instructions in the section Care with KaVo Spray Setup methods according to DIN EN ISO 17664 48 7 7 5 2 C Care products and sys...

Page 14: ...KaVo care products Setup methods according to DIN EN ISO 17664 52 CAUTION Contact corrosion due to moisture Damage to product Immediately remove the product from the steam steriliser after the steril...

Page 15: ...to this KaVo medical device KaVo provides the end customer with a warranty of proper function and guarantees zero defects in respect of material and processing for a period of 24 months from data of i...

Page 16: ...consequences are derived from manipulations or changes to the product by the customer or a third party Service warranty claims will only be accepted if the product is submitted along with proof of pu...

Reviews: