CAUTION
Risk of burn injury from hot handpiece head or hot handpiece lid.
Burn injuries in the mouth may be caused if the instrument overheats.
▶ Never touch soft tissue with the handpiece head or handpiece lid!
CAUTION
Risk due to incorrectly stored instrument.
Damage to clamping system from dropping the instrument.
▶ After treatment, place the instrument properly in the cradle without
the tool.
Safety
11
CAUTION
Instructions for the safe use of handpieces with electronic micromotors.
Electronic micromotors generate much more energy than conventional
pneumatic turbines and motors. Given the higher torque and speed, hand‐
pieces that are poorly serviced, damaged or used improperly can overheat
which can seriously burn the patient.
▶ Never contact soft tissue with the instrument head.
Note
For safety reasons, we recommend that the tool holder system be checked
annually after the warranty period expires.
Safety
12
The following individuals are authorized to repair and service KaVo prod‐
ucts:
▪ Technicians at KaVo branches throughout the world
▪ Technicians specially trained by KaVo
The following guidelines must be observed to ensure save use of the elec‐
trically driven handpieces:
▪ The service instructions for handpieces must be precisely following
when using KAVO spray or QUATTROcare care systems.
▪ Before each use, the handpiece must be checked for external damage.
▪ Before each use, perform a test run with the handpiece, and watch for
atypical heating and unusual noise and vibration.
▪ Immediately stop using handpieces that act unusual.
▪ Never press the pushbutton during operation. This also includes lifting
the cheek or tongue!
Safety
13
To ensure proper function, the medical device must be set up according to
the reprocessing methods described in the KaVo Instructions for Use, and
the care products and care systems described therein must be used. KaVo
recommends specifying a service interval at the dental office for a licensed
shop to clean, service and check the functioning of the medical device. This
service interval depends on the frequency of use and should be adjusted
accordingly.
Service may only be carried out by KaVo-trained repair shops using original
KaVo replacement parts.