77 S
Setup methods according to DIN EN ISO 17664
7.1 P
Preparation at the site of use
WARNING
Hazard from non-sterile products.
There is a risk of infection from contaminated medical devices.
▶ Observe suitable personal protective measures.
▶
Remove all residual cement, composite or blood without delay.
▶
Recondition the medical device as soon as possible after treatment.
▶
Remove the cutter or grinder from the medical device.
▶
The medical device must be dry when transported for reconditioning.
▶
Do not place it in a solution or similar.
Setup methods according to DIN EN ISO 17664
35
77.2 C
Cleaning
CAUTION
Malfunctions from cleaning in an ultrasonic unit.
Defects in the product.
▶ Only clean manually or in a thermodisinfector.
7.2.1 C
Cleaning: Manual cleaning - external
Accessories required:
▪ Tap water 30 °C ± 5 °C (86 °F ± 10 °F)
▪ Brush, e.g. medium-hard toothbrush
Setup methods according to DIN EN ISO 17664
36
▶
Brush off under flowing tap water.
7.2.2 C
Cleaning: Automated external cleaning
KaVo recommends thermodisinfectors in accordance with EN ISO 15883
that are operated with alkaline cleaning agents at a pH of max. 10 (e.g.
Miele G 7781 / G 7881 – validation was performed with the ""VARIO-TD"
programme, "neodisher
®
mediclean" cleaning agent, "neodisher
®
Z" neu‐
tralisation agent, and "neodisher
®
mielclear" rinsing agent and extends only
to the compatibility of materials with respect to KaVo products).
Setup methods according to DIN EN ISO 17664
37
▶
For program settings as well as cleansers and disinfectants to be used,
please refer to the Instructions for Use of the thermodisinfector (com‐
plying with max. pH value of 10).
▶
In order to prevent negative effects on the medical device, make sure
that the interior and the exterior of the medical device are dry, and then
lubricate immediately with care agents from the KaVo care system.
77.2.3 C
Cleaning: Manual cleaning - internal
Can only be done with KaVo CLEANspray or KaVo DRYspray.
▶
Cover the medical device with the KaVo CLEANpac bag, and place it
on the corresponding care adapter. Press the spray button three times
for 2 seconds each time. Remove the medical device from the spray
attachment and let the cleaner work for one minute.
Setup methods according to DIN EN ISO 17664
38