66   TTroubleshooting

6.1   C

Check for malfunctions

CAUTION

Missing or damaged O-rings.

Malfunctions and premature failure.

▶ Make sure that all O-rings are on the coupling and undamaged.

CAUTION

Heating of the product.

Burns or product damage from overheating.

▶ Do not use the product if it is irregularly heated.

 

 

Troubleshooting

31

The medical device is too hot while idling:

Check the amount of cooling air.

The medical device is too hot while working:

Caring for the medical device.

When the speed drops or is uneven:

Caring for the medical device.

An O-ring is missing on the motor coupling:

Replace O-ring.

66.2.1   TTroubleshooting: Exchanging the O-rings on the motor coupling

CAUTION

Hazard from improper care of the O-rings.

Malfunctions or complete failure of the product.

▶ Do not use Vaseline or other grease or oil.

 

 

 

 

 

Troubleshooting

32

Note
The O-rings on the motor coupling may only be lubricated with a cotton

ball wetted with KaVo spray.

Press the O-ring between your fingers to form a loop.

Shove the O-ring to the front, and remove it.

Insert new O-rings into the grooves.

 

 

 

Troubleshooting

33

66.2.2   TTroubleshooting: Cleaning the spray nozzle

CAUTION

Hazard from insufficient amount of spray water.

Insufficient spray water can cause the medical device to overheat and

damage the tooth.

▶ Check spray water channels and if necessary clean spray nozzles

with the nozzle needle (M

Mat. no. 0.410.0921).

 

Troubleshooting

34

Summary of Contents for INTRAmatic contra-angle 80 ES

Page 1: ...ger levels 8 2 2 Safety instructions 9 3 Product description 15 3 1 Purpose Proper use 16 3 2 Technical data 18 3 3 Transportation and storage conditions 19 Contents 1 4 4 First use 21 4 1 Check the a...

Page 2: ...Sterilisation 52 7 8 Storage 55 8 8 Tools and consumables 56 9 Terms and conditions of warranty 58 Contents 4 1 1 U User instructions Dear User Congratulations on purchasing this KaVo quality product...

Page 3: ...NING WARNING indicates a hazardous situation that can cause death or serious injury Safety 8 DANGER DANGER indicates a hazardous situation that can directly cause death or serious injury 2 2 S Safety...

Page 4: ...s at KaVo branches throughout the world Technicians specially trained by KaVo The following guidelines must be observed to ensure save use of the elec trically driven handpieces The service instructio...

Page 5: ...l device may only be used for the described application by a knowledgeable user The following must be observed the applicable health and safety regulations the applicable accident prevention regulatio...

Page 6: ...RH to 95 RH Product description 19 Air pressure 700 hPa to 1060 hPa 10 psi to 15 psi Protect from moisture Product description 20 4 4 F First use WARNING Hazard from nonsterile products Infection dan...

Page 7: ...eration 24 5 5 1 A Attaching the medical device Place the medical device on the LUX motor coupling and lock it into place Before each treatment pull on the medical device to see if it is securely seat...

Page 8: ...r or grinder is securely attached by pulling on it Operation 28 5 5 4 R Removing the milling tool or diamond grinder CAUTION Hazard from rotating tools Laceration Do not carelessly contact rotating to...

Page 9: ...1 T Troubleshooting Exchanging the O rings on the motor coupling CAUTION Hazard from improper care of the O rings Malfunctions or complete failure of the product Do not use Vaseline or other grease o...

Page 10: ...fectors in accordance with EN ISO 15883 that are operated with alkaline cleaning agents at a pH of max 10 e g Miele G 7781 G 7881 validation was performed with the VARIO TD programme neodisher medicle...

Page 11: ...ing agent and extends only to the compatibility of materials with respect to KaVo products For program settings as well as cleansers and disinfectants to be used please refer to the Instructions for U...

Page 12: ...Z neu tralisation agent and neodisher mielclear rinsing agent and extends only to the compatibility of materials with respect to KaVo products For program settings as well as cleansers and disinfecta...

Page 13: ...le tip in the opening and spray Carry out the servicing according to the instructions in the section Care with KaVo Spray Setup methods according to DIN EN ISO 17664 48 7 7 5 2 C Care products and sys...

Page 14: ...KaVo care products Setup methods according to DIN EN ISO 17664 52 CAUTION Contact corrosion due to moisture Damage to product Immediately remove the product from the steam steriliser after the steril...

Page 15: ...to this KaVo medical device KaVo provides the end customer with a warranty of proper function and guarantees zero defects in respect of material and processing for a period of 24 months from data of i...

Page 16: ...consequences are derived from manipulations or changes to the product by the customer or a third party Service warranty claims will only be accepted if the product is submitted along with proof of pu...

Reviews: