Spray the disinfectant on a cloth, then thoroughly wipe down the med‐
ical device and leave the disinfectant to soak in according to the in‐
structions from the disinfectant manufacturer.
▶
Follow the instructions for use of the disinfectant.
77.3.2 D
Disinfection: Manual disinfection - internal
The efficacy of manual internal disinfection must be demonstrated by the
manufacturer of the disinfection agent. With KaVo products, use only dis‐
infection agents that have been released by KaVo with respect to the com‐
patibility of materials (e.g. WL-cid / made by ALPRO).
▶
Immediately after internal disinfection, lubricate the KaVo medical de‐
vice immediately with care agents from the KaVo care system.
Setup methods according to DIN EN ISO 17664
43
77.3.3 D
Disinfection: Machine disinfection - external and internal
KaVo recommends thermodisinfectors in accordance with EN ISO 15883
that are operated with alkaline cleaning agents at a pH of max. 10 (e.g.
Miele G 7781 / G 7881 – validation was performed with the ""VARIO-TD"
programme, "neodisher
®
mediclean" cleaning agent, "neodisher
®
Z" neu‐
tralisation agent, and "neodisher
®
mielclear" rinsing agent and extends only
to the compatibility of materials with respect to KaVo products).
▶
For program settings as well as cleansers and disinfectants to be used,
please refer to the Instructions for Use of the thermodisinfector (com‐
plying with max. pH value of 10).
▶
In order to prevent negative effects on the medical device, make sure
that the interior and the exterior of the medical device are dry, and then
lubricate immediately with care agents from the KaVo care system.
Setup methods according to DIN EN ISO 17664
44
77.4 D
Drying
Manual drying
▶
Blow off the outside and inside with compressed air until water drops
are no longer visible.
Machine drying
The drying procedure is normally part of the cleaning program of the ther‐
modisinfector.
▶
Follow the instructions for use of the thermodisinfector.
Setup methods according to DIN EN ISO 17664
45
77.5 C
Care products and systems - Servicing
WARNING
Sharp cutters or grinders in the medical device.
Risk of injury from sharp or pointed cutters or grinders.
▶ Remove cutter or grinder.
CAUTION
Premature wear and malfunctions from improper servicing and care.
Reduced product life.
▶ Perform proper care regularly!
Setup methods according to DIN EN ISO 17664
46