Product description

INTRAmatic contra-angle handpiece 80 ES,M

Mat. no. 1.004.4375

 

 

Product description

15

33.1   P

Purpose – Proper use

Purpose:

This medical device is

▪ Only intended for dental treatment. Any other type of use or alteration

to the product is impermissible and can be hazardous. The medical

device is intended for the following uses: Removal of carious material,

cavities and crown preparations, removal of fillings, processing of tooth

and restoration surfaces.

▪ A medical device according to relevant national statutory regulations.

 

 

 

Product description

16

P

Proper use:

According to these regulations, this medical device may only be used for

the described application by a knowledgeable user. The following must be

observed:

▪ the applicable health and safety regulations

▪ the applicable accident prevention regulations

▪ these Instructions for use

According to these regulations, the user is required to:

▪ only use equipment that is operating correctly,

▪ use the equipment for the proper purpose,

▪ protect him or herself, the patient and third parties from danger, and

▪ avoid contamination from the product..

 

 

Product description

17

33.2   TTechnical data

Drive speed

max. 40,000 rpm

Speed transmission

1:1

ID

1 blue ring

Pushbutton chuck

1.6 dia.

Milling cutters or grinders

ISO 1797-1 type 3

The contra-angle handpiece can be mounted on all INTRAmatic motors,

and motors with a connection in accordance with ISO 3964 / DIN 13940.

 

 

 

Product description

18

Summary of Contents for INTRAmatic contra-angle 80 ES

Page 1: ...ger levels 8 2 2 Safety instructions 9 3 Product description 15 3 1 Purpose Proper use 16 3 2 Technical data 18 3 3 Transportation and storage conditions 19 Contents 1 4 4 First use 21 4 1 Check the a...

Page 2: ...Sterilisation 52 7 8 Storage 55 8 8 Tools and consumables 56 9 Terms and conditions of warranty 58 Contents 4 1 1 U User instructions Dear User Congratulations on purchasing this KaVo quality product...

Page 3: ...NING WARNING indicates a hazardous situation that can cause death or serious injury Safety 8 DANGER DANGER indicates a hazardous situation that can directly cause death or serious injury 2 2 S Safety...

Page 4: ...s at KaVo branches throughout the world Technicians specially trained by KaVo The following guidelines must be observed to ensure save use of the elec trically driven handpieces The service instructio...

Page 5: ...l device may only be used for the described application by a knowledgeable user The following must be observed the applicable health and safety regulations the applicable accident prevention regulatio...

Page 6: ...RH to 95 RH Product description 19 Air pressure 700 hPa to 1060 hPa 10 psi to 15 psi Protect from moisture Product description 20 4 4 F First use WARNING Hazard from nonsterile products Infection dan...

Page 7: ...eration 24 5 5 1 A Attaching the medical device Place the medical device on the LUX motor coupling and lock it into place Before each treatment pull on the medical device to see if it is securely seat...

Page 8: ...r or grinder is securely attached by pulling on it Operation 28 5 5 4 R Removing the milling tool or diamond grinder CAUTION Hazard from rotating tools Laceration Do not carelessly contact rotating to...

Page 9: ...1 T Troubleshooting Exchanging the O rings on the motor coupling CAUTION Hazard from improper care of the O rings Malfunctions or complete failure of the product Do not use Vaseline or other grease o...

Page 10: ...fectors in accordance with EN ISO 15883 that are operated with alkaline cleaning agents at a pH of max 10 e g Miele G 7781 G 7881 validation was performed with the VARIO TD programme neodisher medicle...

Page 11: ...ing agent and extends only to the compatibility of materials with respect to KaVo products For program settings as well as cleansers and disinfectants to be used please refer to the Instructions for U...

Page 12: ...Z neu tralisation agent and neodisher mielclear rinsing agent and extends only to the compatibility of materials with respect to KaVo products For program settings as well as cleansers and disinfecta...

Page 13: ...le tip in the opening and spray Carry out the servicing according to the instructions in the section Care with KaVo Spray Setup methods according to DIN EN ISO 17664 48 7 7 5 2 C Care products and sys...

Page 14: ...KaVo care products Setup methods according to DIN EN ISO 17664 52 CAUTION Contact corrosion due to moisture Damage to product Immediately remove the product from the steam steriliser after the steril...

Page 15: ...to this KaVo medical device KaVo provides the end customer with a warranty of proper function and guarantees zero defects in respect of material and processing for a period of 24 months from data of i...

Page 16: ...consequences are derived from manipulations or changes to the product by the customer or a third party Service warranty claims will only be accepted if the product is submitted along with proof of pu...

Reviews: