77.6 P
Packaging
Note
The sterilisation bag must be large enough for the instrument so that the
bag is not stretched.
The quality and use of the sterilised product packaging must satisfy ap‐
plicable standards and be suitable for the sterilisation procedure.
▶
Seal each medical device individually in a sterilised item package!
Setup methods according to DIN EN ISO 17664
51
77.7 S
Sterilisation
Sterilisation in a steam steriliser (autoclave) in accordance with EN 13060/
ISO 17665-1 (e.g. KaVo STERIclave B 2200 / 2200 P)
CAUTION
Premature wear and malfunctions from improper servicing and care.
Reduced product life.
▶ Before each sterilisation cycle, service the medical device with KaVo
care products.
Setup methods according to DIN EN ISO 17664
52
CAUTION
Contact corrosion due to moisture.
Damage to product.
▶ Immediately remove the product from the steam steriliser after the
sterilisation cycle!
The KaVo medical device has a maximum temperature resistance up to
138 ℃ (280.4 °F).
Setup methods according to DIN EN ISO 17664
53
(Depending on the available autoclave,) select a suitable procedure from
the following sterilisation processes:
▪ Autoclave with three times initial vacuum:
–
at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4
°F)
Drying time: 20 min.
▪ Autoclave using the gravitation method:
–
at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4
°F) or alternatively
Drying time: 30 min.
–
at least 60 minutes at 121 °C -1 °C / +4 °C (250 °F -1.6 °F / +7.4
°F)
Drying time: 30 min.
▶
Use according to the manufacturer's Instructions for Use.
Setup methods according to DIN EN ISO 17664
54