Care instructions ESTETICA E30
2 Reprocessing methods according to DIN EN ISO 17664 | 2.2 General instructions for the reprocessing of KaVo
treatment centres
Manual drying
▶
Supply dry, clean and uncontaminated compressed air according
to ISO 7494-2 only.
▶
Blow off the outside and inside with compressed air until water
drops are no longer visible.
Automatic drying
Usually, automatic drying is part of the cleaning/disinfection cycle of
the washer disinfector and should not be carried out above 95 °C.
Servicing
▶
Discard damaged products.
Control and functional test
▶
Perform a visual inspection to check for damage and wear and
tear.
▶
Test products for function regularly.
Packaging
▶
Use standardised packaging material. Sealing should not be un-
der tension.
▶
Use sterile goods packaging in order to ensure sterility.
Sterilisation
All sterilisable products are temperature-resistant up to max. 138
°C.
KaVo recommends sterilisation in a steam steriliser (autoclave) ac-
cording to EN 13060 / ISO 17665-1, e.g. STERIclave B 2200 / 2200
P (KaVo) or Citomat K-Series (Getinge).
▶
Sterilise with a triple fractionated vacuum (B class steriliser);
hold for 4 minutes; 134°C ± 1 °C; overpressure: 2.13 bar.
Storage
▶
Reprocessed products should be stored protected from dust with
minimum exposure to germs in a dry, dark and cool place.
▶
Comply with the expiry date of the sterilised items.
2.2.1 Overview ESTETICA E30 reprocessing
Reprocessing items/medical
devices
Cleaning
Disinfection
Sterilisa-
tion
manual
automatic manual
automatic automatic
Surfaces
Device
X
X
Cushion
X
X
Light
X
X
See also:
2
IfU KaVoLUX 540
LED / EDI / MAIA
Lights, handles
X
X
X
X
X
Dentist element
Handle TM table / handle cup S table
X
X
Cannula triple-function handpiece
X
X
X
Holder pad/anti-slip mat
X
X
X
X
X
Handpiece holder
X
X
Instrument arm S table
X
X
Assistant element
Connection parts of the suction hose
X
X
X
X
Patient element
Tumbler holder
X
X
Tumbler filler
X
X
Porcelain spittoon bowl
X
X
X
X
7 / 40