Care instructions ESTETICA E30
2 Reprocessing methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist element
26 / 40
Note
Handpieces must be sterilised before each use. Non-sterile handpieces and
tips can cause bacterial or viral infections.
Note
Please comply with the current local regulations governing the re-use and
disposal of equipment.
The medical device is resistant to temperatures of up to 138
℃
(280.4
o
F).
▶
Place the cleaned and disinfected handpieces separately in sterilisation
packages and weld them to be sealed, e.g. KaVo STERIclave bags
Mat. no.
0.411.9912
).
135°C
KaVo recommends, e.g.
- STERIclave B 2200/ 2200P from KaVo
Citomat / K series from Getinge
Autoclave with a triple pre-vacuum for at least 4 minutes at 134 °C ± 1 °C
(273.2 °F ± 33.8 °F)
For range of applications, refer to the manufacturer's Instructions for Use.
Only for handpieces with fibre optic conductor sleeve
▶
If the fibre optic conductor sleeve
Mat. no. 1.007.4021
looses its bright-
ness due to sterilisation, replace the fibre optic conductor sleeve.
The light source in the handpiece cannot be replaced.
Sterilisation of tips, endo files, file holders, endo
wrenches, torque wrenches
Note
The maximal number of sterilisation cycles must not be exceeded.