Care instructions ESTETICA E30
2 Reprocessing methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist element
NOTICE
Damage caused by using a nozzle needle that is too long.
Damage to the internal hoses of the cannula.
▶
Use a short nozzle needle (
Mat. no. 1.004.4986
) to clean the media exit
holes.
▶
Carefully remove contamination from the media exit ports using the short
nozzle cleaning needle.
Disinfection of triple-function handpiece KaVo One
Manual disinfection
See also:
2
2.2 General instructions for the reprocessing of KaVo treatment centres,
Page 6
External disinfection
▶
Wipe-disinfect the surfaces with a soft cloth.
Automatic disinfection
NOTICE
Damage caused by improper cleaning/disinfection.
▶
Do not sterilise the single parts of the cannula in a hot-air steriliser.
▶
Do not place the single parts of the cannula in a disinfectant or ultrasonic
bath.
▶
Single parts of the cannula should be subjected to thermodisinfection and
sterilisation.
Drying of triple-function handpiece KaVo One
Manual drying
▶
Blow off the outside and inside with compressed air until water drops are no
longer visible.
Automatic drying
Not applicable.
Sterilisation of triple-function handpiece KaVo One
Note
The cannula can be subjected to sterilisation.
▶
Hold the valve body of the triple-function handpiece and screw off the can-
nula.
13 / 40