Care instructions ESTETICA E30

2 Reprocessing methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist element

NOTICE

Damage caused by improper cleaning/disinfection.

Do not wash the gripping sleeve and the cannula in the washer disinfector.

Do not sterilise the gripping sleeve and the cannula in the hot-air steriliser.

Do not place the gripping sleeve and the cannula in a disinfectant or ultra-
sonic bath.

Manual cleaning

Note

Combine manual cleaning with disinfection.

Internal cleaning

Leave the gripping sleeve and cannula on the instrument, and activate the
air and water for 60 seconds.

External cleaning

 CAUTION

Danger of injury if the valve body is hot (heating elements, high-pres-
sure lamp).

Risk of burn injury.

Switch off the unit at the main switch.

Let the handpiece cool after long use.

Accessories required:

▪ Tap water 30

o

C ± 5

o

C

▪ Brush, e.g. medium-hard toothbrush

Disassemble the gripping sleeve and cannula.

Detach the key part from the gripping sleeve.

 

Clean the key part, gripping sleeve and cannula by brushing them under
running water (at least tap water quality).

Immediately remove contamination from materials used in the dental prac-
tice (impression materials, caustic chemicals).

Carefully wipe off contamination from the cannula tip with a soft Q-tip or
soft cloth and water.

NOTICE

Damage caused by using a nozzle needle that is too long.

Damage to the internal hoses of the cannula.

Use a short nozzle needle (

Mat. no. 1.004.4986

) to clean the media exit

holes.

15 / 40

Summary of Contents for ESTETICA E30

Page 1: ...Care instructions ESTETICA E30 ...

Page 2: ...Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 ...

Page 3: ...oses 17 2 4 7 Servicing the turbine return air filter 18 2 4 8 Reprocessing KaVo ERGOcam One 18 2 4 9 Reprocessing KaVo PiezoLED 20 2 5 Reprocessing of the assistant element 27 2 5 1 Preparing the handpiece holder 27 2 5 2 Reprocessing of the suction hose connector 28 2 5 3 Reprocessing of sieve inserts and suction hoses 29 2 6 Reprocessing of the patient element 29 2 6 1 Reprocessing of the tumbl...

Page 4: ... instructions TI Technician s instructions SC Safety checks IEC International Electrotechnical Commission RI Repair instructions EMC Electromagnetic compatibility Scaler Scaler ZEG IR Infrared IrDA Infrared Data Association 1 1 2 Symbols Refer to the chapter on Safety Warning symbol Important information for users and service technicians Action request CE mark according to Medical Devices Directiv...

Page 5: ...ng of the treatment unit Lack of instructions to the cleaning staff and lack of preparation of the treat ment unit can lead to the cleaning personnel sustaining injuries Only trained professionals and instructed cleaning personnel may be present in the treatment rooms Position the chair for cleaning and turn the device off CAUTION Improper disinfection may lead to health hazards and material dam a...

Page 6: ...to the Instructions for Use of the washer disinfector Directly after automated cleaning disinfection treat the medical device with the care products and systems provided by KaVo Manual disinfection Perform wipe disinfection only the use of spray disinfection and disinfection baths is not feasible Wipe all outer surfaces with a soft cloth and an approved disin fectant Approved disinfectants CaviCid...

Page 7: ...nt up to max 138 C KaVo recommends sterilisation in a steam steriliser autoclave ac cording to EN 13060 ISO 17665 1 e g STERIclave B 2200 2200 P KaVo or Citomat K Series Getinge Sterilise with a triple fractionated vacuum B class steriliser hold for 4 minutes 134 C 1 C overpressure 2 13 bar Storage Reprocessed products should be stored protected from dust with minimum exposure to germs in a dry da...

Page 8: ...with water immediately NOTICE Damage caused by spray disinfection Disinfectant gets into cracks and loosens the paint Do not use spray disinfection Only disinfect by wiping Clean the surfaces with a soft cloth and water and wipe disinfect them with an approved surface disinfectant 2 3 Reprocessing of the surfaces of treatment centre upholstery Position the chair for cleaning Adjust the backrest to...

Page 9: ...rex Microcide AF Liquid Schülke Mayr FD 322 Dürr Dental Incidin Liquid Ecolab Clean the surfaces and upholstery with a soft cloth and water and wipe dis infect them with an approved surface disinfectant NOTICE Damage caused by spray disinfection Disinfectant gets into cracks and loosens the paint Do not use spray disinfection Only disinfect by wiping Cleaning and disinfection of the foot control N...

Page 10: ...d wipe disinfect it with an approved surface disinfectant Cleaning disinfection and sterilisation of holder pad anti slip mat Sterilise the holder pad and anti slip mat by thermodisinfection 2 4 2 Reprocessing of the instrument holder of the TM table Note The handpiece holder can be removed for better cleaning Cleaning and disinfection of the handpiece holder NOTICE Damage caused by treatment in a...

Page 11: ...e cleaning Cleaning and disinfection of the instrument arm S table NOTICE Damage caused by treatment in a washer disinfector Do not sterilise the instrument arm S table in a washer disinfector Take the handpieces off the hoses Take the roller out of the instrument arm Pull off the instrument arm in an upward direction Clean the instrument arm and roller under running water Wipe the instrument arm ...

Page 12: ... the single parts of the cannula in a hot air steriliser Do not place the single parts of the cannula in a disinfectant or ultrasonic bath Single parts of the cannula should be subjected to thermodisinfection and sterilisation Manual cleaning Note Combine manual cleaning with disinfection Internal cleaning Leave the cannula and handpiece on the instrument and activate the air and water for 60 seco...

Page 13: ... External disinfection Wipe disinfect the surfaces with a soft cloth Automatic disinfection NOTICE Damage caused by improper cleaning disinfection Do not sterilise the single parts of the cannula in a hot air steriliser Do not place the single parts of the cannula in a disinfectant or ultrasonic bath Single parts of the cannula should be subjected to thermodisinfection and sterilisation Drying of ...

Page 14: ...the cannula Immediately remove the parts from the steriliser after the sterilisation cycle is completed Storage Reprocessed products should be stored protected from dust with minimum exposure to germs in a dry dark and cool place 2 4 5 Preparation of triple function handpiece Note Wet the O rings of the gripping sleeves after sterilisation using silicone grease Mat no 1 000 6403 or KaVo Rota Spray...

Page 15: ...injury if the valve body is hot heating elements high pres sure lamp Risk of burn injury Switch off the unit at the main switch Let the handpiece cool after long use Accessories required Tap water 30o C 5o C Brush e g medium hard toothbrush Disassemble the gripping sleeve and cannula Detach the key part from the gripping sleeve Clean the key part gripping sleeve and cannula by brushing them under ...

Page 16: ...in the hot air steriliser Do not place the gripping sleeve and the cannula in a disinfectant or ultra sonic bath Drying of triple function handpiece Manual drying Blow off the outside and inside with compressed air until water drops are no longer visible Automatic drying Not applicable Care of triple function handpiece Requirement The multifunctional handpiece has been prepared Apply KaVo silicone...

Page 17: ...rpressure 2 13 bar NOTICE Damage caused by over heating Damage to the gripping sleeve Immediately remove the parts from the steriliser after the sterilisation cycle is completed Storage Reprocessed products should be stored protected from dust with minimum exposure to germs in a dry dark and cool place 2 4 6 Preparation of the instrument hoses Cleaning and disinfection of handpiece hoses Clean the...

Page 18: ...ctor must be checked weekly If there is any liquid in the filter drain the liquid and replace the O rings of the multiflex couplings If necessary replace the turbine return air filter Mat no 02003098 and clean the filter holder Note The turbine return air filter is in the same position in the dentist element vari ant with S table 2 4 8 Reprocessing KaVo ERGOcam One Note The camera handle cannot be...

Page 19: ...ra Clean the surface with a soft cloth and mild liquid cleaning agent Disinfection Exterior disinfection by hand WARNING Highly flammable gases Explosion Allow product to cool before disinfecting NOTICE Damage to the camera due to improper disinfection Irreversible destruction of the camera Disinfect with wipes only do not use disinfectant sprays Do not immerse the camera in liquids Do not sterili...

Page 20: ...ly with the manufacturer s instructions for use with regard to the applic ation range and exposure time NOTICE Material damage due to use of non approved disinfectant Damage to the camera Use approved disinfectants only 2 4 9 Reprocessing KaVo PiezoLED Cleaning NOTICE Malfunctions from cleaning in the ultrasonic unit Defects in the device Only clean manually or in a washer disinfector Only product...

Page 21: ...e to the back nozzle Rinse in the normal flow direction do not rinse against the flow direction If an aldehyde free cleaning and disinfection solution is used subsequently rinse at least thrice with distilled or deionised water Automated external cleaning KaVo recommends washer disinfectors according to EN ISO 15883 1 which are operated using alkaline cleaning agents having a maximum pH value of 1...

Page 22: ... if the product continues to contain visible contamination the cleaning process must be repeated Disinfection NOTICE Malfunctioning from using a disinfectant bath or chloride containing disinfectants Defects in the device Disinfect manually only Disinfection of handpieces Note For times and concentrations please refer to the instructions of the manufac turers of the cleaning disinfection agent Man...

Page 23: ...on of tips endo files file holders endo wrenches torque wrenches Place the products in the cleaning solution at least for as long as specified by the manufacturer of the cleaning disinfection agent Remove all contamination from the outer surface by careful brushing using a soft brush or soft cloth Rinse the inside of products thoroughly at least five times using fresh dis tilled or deionised water...

Page 24: ...he products with filtered compressed air max 3 bar If required dry again at a clean place Package products immediately after drying see section on Packaging and Sterilisation Servicing CAUTION Check handpiece and hose for visible damage prior to use They need to be replaced if any damage is evident CAUTION Check tips for visible damage and wear and tear prior to use If the damage or wear and tear ...

Page 25: ...ackaging Note The sterilisation bag must be large enough for the instrument so that the bag is not stretched The quality and use of the sterilisation packaging must comply with applic able standards and be suitable for the sterilisation procedure The medical device must be packed before sterilisation Sterilisation Sterilisation of handpieces Sterilisation in a steam steriliser Autoclave EN 13060 I...

Page 26: ...nd weld them to be sealed e g KaVo STERIclave bags Mat no 0 411 9912 135 C KaVo recommends e g STERIclave B 2200 2200P from KaVo Citomat K series from Getinge Autoclave with a triple pre vacuum for at least 4 minutes at 134 C 1 C 273 2 F 33 8 F For range of applications refer to the manufacturer s Instructions for Use Only for handpieces with fibre optic conductor sleeve If the fibre optic conduct...

Page 27: ...erilisation cassette e g KaVo sterilisation cassette Mat no 0 411 9101 Sterilisation container requirements EN 868 and ISO 11607 Resistant up to 138 o C with appropriate permeability for steam Regular servicing The requirements also apply to double disposable sterilisation packages Permissible sterilisation apparatus Sterilisation apparatus with validated cycle parameters Sterilisation apparatus w...

Page 28: ...ng and disinfection of the suction hose connector Clean the disassembled parts under running water Thermodisinfect the disassembled connectors and for at least 10 minutes at 93 C total cycle approximately 20 minutes Disinfect the connectors and by wipe disinfection Note Connectors and cannot be subjected to thermodisinfection Sterilising the suction hose connector NOTICE Damage caused by sterilisa...

Page 29: ...automatically when the suction tubes are removed from the holder Pull the sieve inserts out of the housing Place a cloth under the sieve seat If needed insert a new sieve insert Mat no 10088636 into the seat Wipe the suction tubes with disinfectant 2 6 Reprocessing of the patient element 2 6 1 Reprocessing of the tumbler Reprocess as specified by the manufacturer 2 6 2 Reprocessing of tumbler hold...

Page 30: ... disinfector Sterilisation of tumbler holder and tumbler filler NOTICE Damage caused by sterilisation Do not sterilise the tumbler holder and filler 2 6 3 Reprocessing of the spittoon bowl NOTICE Water in the device Material damage from leaking water upon actuation of the Tumbler filler or Bowl flush keys Make sure that the unit is turned off before removing the tumbler filler or the spittoon bowl...

Page 31: ...ive germ reduction Note KaVo recommends rinsing all water conducting systems at all dispensing sites while no handpiece is attached such as e g syringe turbine motor ultrasonic scaler and tumbler filler at the start of every working day for 2 minutes after every patient for 20 seconds Note The continuous germ reduction hygiene function is available only with the water bottle kit Treatment water th...

Page 32: ... must be carried out At first start up Before during and after prolonged downtimes more than 4 weeks Over the weekend During vacations After treatment of patients at high risk Order of manual intensive germ reduction Fill the intensive germ reduction bottle See also 2 3 1 1 Filling the intensive germ reduction bottle Page 37 Press the button during the rinse Rinse each dental instrument for 20 sec...

Page 33: ...e Regarding the air jet suction system Venturi only one aliquot may be drawn into each hose Aspirate one tumbler of cold water with each suction hose With the suction handpiece attached or by hand press down the DEKA SEPTOL dosing facility once each for the saliva ejector hose and the spray mist ejector Aspirate the DEKASEPTOL Gel from the dosing facility Note In suction processes with air jet suc...

Page 34: ...Care instructions ESTETICA E30 2 Reprocessing methods according to DIN EN ISO 17664 2 8 Reprocessing of the suction system 34 40 ...

Page 35: ...if it is scratched deformed or discoloured Note the maximum useful life of the water bottle on the label Water bottle complete Mat no 0 710 4151 Dosing attachment Mat no 1 002 0287 Nameplate of water bottle and intensive germ reduction bottle 3 2 1 4 7 5 6 8 3 2 1 4 7 5 6 8 Nameplate of water bottle intensive germ reduction bottle Manufacturer Concentration Use by date Material number Month manufa...

Page 36: ...nsure the efficiency of the measures KaVo urgently recommends repro cessing the water with KaVo OXYGENAL 6 For cleaning rinse the water bottle with water at less than 50 o C Filling the water bottle with the dosing facility Wipe up any spilled OXYGENAL immediately in order to prevent discoloura tion of the surfaces Remove the protective hose from the tube of the KaVo OXYGENAL 6 dosing facility Gui...

Page 37: ...rm reduction liquid or the dose is too high Injury hazard Use KaVo OXYGENAL 6 only Do not exceed the maximum concentration of 0 25 KaVo OXYGENAL 6 Proceed like for filling the water bottle but use a different dose of KaVo OXYGENAL 6 See also 2 Filling the water bottle with the dosing facility Page 36 Fill the four fold dose of KaVo OXYGENAL 6 equivalent to 20 ml in the in tensive germ reduction bo...

Page 38: ...solve after several dosing processes Screw the intensive germ reduction bottle tightly to the treatment centre through a clockwise rotation ð Apply compressed air to the intensive germ reduction bottle 3 2 Check the OXYGENAL 6 concentration Fill the tumbler with water Use Merckoquant peroxide test strips 100 to 1000 mg l H2O2 Mat no 0 229 5003 to determine the OXYGENAL 6 concentration in the water...

Page 39: ......

Page 40: ...1 009 7934 Fk 20171018 5 en ...

Reviews: