Care instructions ESTETICA E30
2 Reprocessing methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist element
16 / 40
▶
Carefully remove contamination from the media exit ports using the short
nozzle cleaning needle.
Disinfection of triple-function handpiece
Manual disinfection
External disinfection
▶
Wipe-disinfect the surfaces with a soft cloth.
Automatic disinfection
Not applicable.
NOTICE
Damage caused by improper cleaning/disinfection.
▶
Do not wash the gripping sleeve and the cannula in the washer disinfector.
▶
Do not sterilise the gripping sleeve and the cannula in the hot-air steriliser.
▶
Do not place the gripping sleeve and the cannula in a disinfectant or ultra-
sonic bath.
Drying of triple-function handpiece
Manual drying
▶
Blow off the outside and inside with compressed air until water drops are no
longer visible.
Automatic drying
Not applicable.
Care of triple function handpiece
Requirement
The multifunctional handpiece has been prepared.
▶
Apply KaVo silicone grease (
Mat. no. 10006403
) or KaVo Rota Spray 2
(
Mat. no. 04117510
) to the O-rings of the joint between the gripping
sleeve and cannula. Use cotton buds for this purpose.
Sterilisation of triple-function handpiece
Note
The grip sleeve and cannula can be sterilised.
▶
Hold the 3-way or multifunctional handpiece at the gripping sleeve and take
off the cannula with a slight twisting motion.