Care instructions ESTETICA E30
2 Reprocessing methods according to DIN EN ISO 17664 | 2.5 Reprocessing of the assistant element
Note
The use of hot-air sterilisation and radio-sterilisation is not permissible
(causes destruction of the products). KaVo shall not be held responsible if
non-permissible procedures such as ethylene oxide, formaldehyde, and low
temperature plasma sterilisation are used.
Note
Only cleaned and disinfected products may be sterilised.
Note
Please comply with the current local regulations governing the re-use and
disposal of equipment.
▶
Place the cleaned and disinfected tips, endo files, file holders, endo
wrenches and torque wrenches separately in sterilisation packages (e.g.
KaVo STERIclave bags
Mat. no. 0.411.9912
) and weld them to be sealed
or sterilise them in a sterilisation cassette (e.g. KaVo sterilisation cassette
Mat. no. 0.411.9101
).
Sterilisation container requirements:
▪ EN 868 and ISO 11607
▪ Resistant up to 138
o
C with appropriate permeability for steam
▪ Regular servicing
The requirements also apply to double disposable sterilisation packages.
Permissible sterilisation apparatus:
▪ Sterilisation apparatus with validated cycle parameters
▪ Sterilisation apparatus with non-validated cycle parameter which comply
with DIN EN ISO 14161:2000
Permissible procedures:
Procedures
Time /Temperature
Fractionated pre-vacuum
3 to 20 minutes at 132 °C/ 134 °C
Steam sterilisation apparatus
(AAMI TIR no. 12, DIN EN ISO 14161,
DIN EN ISO 17665)
(DQ, IQ, OQ and PQ)
138
o
C
Storage
▶
Reprocessed products should be stored protected from dust with minimum
exposure to germs in a dry, dark and cool place.
▶
Comply with the expiry date of the sterilised items.
2.5 Reprocessing of the assistant element
2.5.1 Preparing the handpiece holder
Note
The handpiece holder can be removed.
27 / 40
Summary of Contents for ESTETICA E30
Page 1: ...Care instructions ESTETICA E30 ...
Page 39: ......
Page 40: ...1 009 7934 Fk 20171018 5 en ...