Care instructions ESTETICA E30
2 Reprocessing methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist element
Cleaning
Note
Disinfect the camera before cleaning.
Disinfect after cleaning if necessary.
Exterior cleaning by hand
NOTICE
Damage to the camera due to improper cleaning.
Destroys the surface or the optical system of the camera.
▶
Do not use abrasives or aggressive cleansers!
▶
Do not immerse the camera in liquids!
▶
Switch main device switch off.
▶
Remove the camera.
▶
Clean the surface with a soft cloth and mild liquid cleaning agent.
Disinfection
Exterior disinfection by hand
WARNING
Highly flammable gases-.
Explosion.
▶
Allow product to cool before disinfecting.
NOTICE
Damage to the camera due to improper disinfection.
Irreversible destruction of the camera.
▶
Disinfect with wipes only, do not use disinfectant sprays!
▶
Do not immerse the camera in liquids!
▶
Do not sterilise the camera in an autoclave!
▶
Do not store the camera in Steri pack!
▶
Do not sterilise the camera with hot air or in a washer disinfector!
▶
Switch main device switch off.
▶
Allow the camera to cool down.
▶
Remove the camera.
▶
Disinfect the surface using a soft wipe and disinfectant.
19 / 40